Quick answer. The thing to understand before buying any portable hardness tester is that none of them measures hardness. A rebound instrument measures the velocity ratio of a small impacting ball; an ultrasonic contact impedance instrument measures the frequency shift of a loaded diamond-tipped rod. Each then converts that measurement into a Brinell, Rockwell or Vickers number through a table that depends on the material group. So a portable reading quoted in HRC is a converted estimate, and it is only valid if the conversion matches the material. Everything else that decides whether you get usable numbers follows from this: the specimen mass and surface finish, the impact direction, the probe type, and a calibrated test block to verify against. Two notes on this listing: roughly half of it tests concrete rather than metal, and several records carry third-party instrument brand names - see the note below before ordering by code.
A bench hardness tester presses a defined indenter with a defined force and measures the indentation - that is a direct measurement, and it is why bench testers remain the reference. A portable instrument cannot carry that load frame into the field, so it measures a dynamic or acoustic proxy instead and converts. Understanding which proxy decides what the instrument can and cannot do.
| Method | What it measures | Suits | Its main limitation |
|---|---|---|---|
| Rebound (dynamic) hardness | The ratio of rebound velocity to impact velocity of a spring-propelled ball | Large, heavy, solid parts; castings, forgings, shafts, weld areas | Needs a minimum specimen mass and thickness, and is sensitive to impact direction. Reads low on anything that can move or ring. |
| Ultrasonic contact impedance | The frequency shift of a vibrating diamond-tipped rod pressed into the surface | Thin walls, small parts, gear teeth, tight spaces, thin coatings on a hard base | Needs a smoother, cleaner surface than rebound, and a steady hand or a stand to apply the load consistently. |
| Bench indentation (for reference) | The indentation itself, under a defined load | Laboratory work and anything that must be reported as a measured value | Not portable. See hardness testing machines. |
The practical rule that follows: if a result has to be defensible as a measured hardness - for a certificate, a dispute, an acceptance test - it should be made on a bench tester. Portable instruments are for screening, for sorting, for checking a heat treatment across a large component, and for the many cases where the part cannot be cut or carried. Both are legitimate; confusing them is not.
Ask which scales the instrument converts to, and which material groups its conversion tables cover - typically steel and cast steel are treated separately from cast iron, stainless, aluminium alloys, brass and bronze. Using a steel table on cast iron produces a confident wrong answer. Ask also whether a user-defined conversion can be entered, because for an unusual alloy the correct procedure is to establish your own correlation by testing reference pieces on a bench tester and on the portable instrument, and an instrument that cannot store that correlation forces the correction to be done by hand every time.
For a rebound instrument this is the specification that is ignored most often and matters most. The method assumes the specimen does not move or ring when struck. Below a minimum mass the part recoils and absorbs energy, and the reading comes out low - sometimes dramatically. Below a minimum thickness the surface flexes, with the same effect. The remedies are standard practice:
Ask for the minimum mass and thickness figures for each probe you are quoted, and check them against the thinnest and lightest part you actually test, not the average.
A portable hardness tester is verified against a reference block of certified hardness, and the block is a consumable: its surface wears where it is struck, so readings must be spread and the block eventually retired. So:
Several records in this category are rebound test hammers for concrete, which despite the shared word "rebound" are not metal hardness testers at all. A concrete test hammer measures the surface rebound of a spring-driven mass against a concrete surface and reports a rebound number, which is then correlated to compressive strength. Three things a buyer must know:
If concrete assessment is your purpose, these belong with concrete testing machines and with core testing - see core cutting and preparation for the coring and specimen preparation that a real correlation requires.
"Ultrasonic instrument" covers two different products that are rarely interchangeable, so confirm which is being quoted:
For either, ask for the probes included, the couplant, the calibration blocks, and the battery and charging arrangement for field use.
1. The category is named for metal and roughly half of it is for concrete. Of the records on the first page, four are portable metal hardness testers, five are concrete rebound test hammers and their calibration anvil, and one is a general ultrasonic instrument. We have recommended that the concrete records move to the concrete and aggregate ranges, where a buyer looking for them would actually search.
2. Several records carry third-party instrument brand names. Where a named branded instrument is supplied, it is supplied as that manufacturer's product rather than as a Ray Export design, and the listing does not currently make that distinction clear. We have raised this internally for correction. In the meantime, please ask us to confirm in writing exactly what is being offered against your enquiry - the manufacturer, the model and whether it is the branded instrument or an equivalent - because you are entitled to know which you are buying and it changes both the price and the support arrangements.
3. One name is truncated mid-word. One record reads "...Portab for engineering schools Hardness Tester", where the word is "Portable" and a suffix has been spliced into the middle of it. It is a data fault rather than a product fault, and it is flagged for correction, but it is worth knowing if you are searching this listing by name and finding nothing.
So enquire by method and application: say whether the material is metal or concrete, the thinnest and lightest part you test, the scale you need to report in, and whether the result must be defensible as a measurement or is for screening. We will quote the right instrument, name the manufacturer, and confirm the conversion tables, probes, mass limits and certificates in writing.
Parent category: non-destructive testing. For laboratory reference testing see hardness testing machines, metal testing machines and universal testing machines; for concrete and cores, concrete testing machines and rock testing and core preparation; for microstructure behind a hardness result, metallurgical equipment. The full TVET training equipment category lists the rest.
No. A rebound instrument measures the velocity ratio of an impacting ball; an ultrasonic contact impedance instrument measures the frequency shift of a loaded diamond-tipped rod. Each converts that measurement into Brinell, Rockwell or Vickers through a material-dependent table. So a portable reading in HRC is a converted estimate, valid only if the conversion matches the material group. For a result that must be defensible as a measurement, use a bench tester.
Because the method assumes the specimen does not move or ring when struck. Below a minimum mass the part recoils and absorbs impact energy; below a minimum thickness the surface flexes. Either way the rebound is reduced and the hardness reads low, sometimes dramatically. Clamp or couple a light part to a support mass with coupling paste, or use the ultrasonic method, which is far less sensitive to specimen mass.
Which scales the instrument converts to and which material groups the tables cover, since steel, cast iron, stainless, aluminium alloys and copper alloys are treated separately and using the wrong table gives a confident wrong answer. Ask also whether a user-defined correlation can be stored, because for an unusual alloy the correct procedure is to establish your own correlation against a bench tester.
Yes, and in a hardness range near your working range - a block at 60 HRC does not verify an instrument used at 150 HB. Ask for the block with its certificate, agree a verification interval, and keep a short log beside the instrument. Note that the block is a consumable: its surface wears where it is struck, so spread the readings and retire it when the surface is used up.
It gives a rebound number, and converting that to strength with the instrument's default curve is an estimate rather than a test result, because the correlation depends on the mix, aggregate, age, moisture and surface carbonation. A defensible figure comes from correlating hammer readings against cores taken from the actual structure. Remember also that the hammer reads only the outer few millimetres, so a hardened or carbonated skin can read high over weaker concrete.
They are different products and rarely interchangeable, so ask. A thickness gauge measures remaining wall thickness, often through paint, and is specified by thickness range and probe frequency. A flaw detector finds and sizes internal defects and needs angle probes, calibration blocks, a documented technique and a qualified operator - so budget for operator qualification, or the instrument will be one nobody is permitted to interpret.
Ray Export is a manufacturer and exporter of laboratory, scientific and educational equipment based at Plot No. 449, HSIIDC Industrial Area, Saha, Ambala, Haryana 133104, India. Enquiries: [email protected], +91-9817319615.
Written by Palak Mall, Senior Content Writer. Technically reviewed by Arvind Kumar, Lab Equipment Specialist, 12+ years in laboratory and testing instrumentation. Last updated: 17 October 2026.
This page describes instrument types generically and quotes no prices; prices depend on method, probes, blocks, certification and destination, and are issued only as dated written quotations. Where a record in this category names a third-party instrument brand, that instrument is the product of its own manufacturer and not a Ray Export design; the manufacturer, model and whether a branded instrument or an equivalent is being offered will be stated in writing on the quotation. No brand relationship is claimed on this page. Conversion scales and material groups, minimum specimen mass and thickness, probe coverage, direction compensation, required surface finish, test block values and certificates, ultrasonic probe frequencies and calibration blocks, battery life and data output will be confirmed in writing before order. No test standard is named on this page and no tolerance or conversion figure is quoted, because the applicable standard and the valid conversion differ by country, instrument and material - name the standard you work to and compliance will be confirmed against it rather than assumed. Statements about method limitations are general practice rather than claims about any particular product. The record counts and the truncated name described above were read from this listing on 16 October 2026 and will change as it is corrected.
1. Good for all metals
2. The deal for production level testing
3. Best suited for on-site testing of heavy, big or already installed parts
4. Handy for difficult to access or confined test locations
5. Automatic compensation for impact direction
6. Excellent for material selection and acceptance tests
7. Easy to use and accurate on curved test surfaces (R>10 mm) Primary Industries: Metal production & Processing Automotive & Transportation Machinery & Power Plants Petro-Chemical, Refineries Aerospace & shipyard Metal Constructions Testing Services & laboratories.
1. Good for all materials 2. Ideal for production level testing 3. Best suited for on-site testing of heavy, big or already installed parts 4. Handy for difficult to access or confined test locations 5. Automatic sensing and compensation for impact direction 6. Excellent for material selection and acceptance tests 7. Easy to use and accurate on curved test surfaces (R>10 mm) Primary Industries: Metal production & processing Automotive & transportation Machinery & power plants Petro-Chemical, refineries Aerospace & shipyard Metal constructions Testing services & laboratories.
1. Ideal for consultants, machinists, quality and production personnel with periodic requirements for quick and easy hardness measurements in HRC 2. Measures 20 - 70 HRC, ± 1 HRC 3. Parts weighing less than 5 kg (11 lbs.) with thin walls must be set on a solid support to prevent vibrations 4. Parts weighing less than 2 kg (4.5 lbs.) must be coupled to a solid support using coupling paste 5. Minimum surface layer thickness: 0.7 mm 6. Minimum test radius of surface curvature: R = 30 mm (1.18 in.) 7. Steel and cast steel, cold work tool steel, stainless steel and high-temperature resistant steel.
Equostat Hardness Tester for NDT Metal Testers - Static Low-Load Measurement based on the Rockwell Principle. Accurate, rapid, and reproducible results from measurement of penetration depth under defined forces and therefore - Independence of material (no choice of material curves necessary) - Independence of body dimensions and geometry - Measures from one side - no backing necessary Excellent sensitivity through small penetration of a few µm and therefore particularly suitable for a. Scratch-sensitive and polished surfaces b. Thin parts, profiles, and pipes with a wall thickness=2 mm c. Hardness profile in the heat-affected zone of welding seams.
SCHMIDT Test Hammer has a Measuring range of 10 to 70 N/mm² compressive strength. Impact energy = 2,207 Nm. Rebound values are recorded as a bar chart on a paper strip. One roll of paper strip offers space for 4000 test impacts. A bridge concreted in several stages is tested for uniform concrete quality. The engineer will perform a series of tests in intervals of 10 m each..
SCHMIDT Test Hammer Type DIGI-SCHMIDT has a Measuring range of 10 (ND) / 18 (LD) to 70 compressive strength. Rebound values are been calculated by an electronic method and may be read directly as compressive strength values..
Silver SCHMIDT Test Hammer - For Extended Range of Applications - Conversion bends are given for an extensive variety of compressive concrete strength, including low and high quality concrete fc<10 N/mm2 (1.450 psi) and up to 170 N/mm2 (24.650 psi). 2. Conversion curves for various sorts of current cement are preset in the Silver Schmidt, in view of tests performed by a free foundation. Operation : 1. Straightforward operation with the \"one button\" UI 2. Dialect autonomous using realistic UI 3. Programmed transformation to the required estimation unit (N/mm2, kg/cm2, psi) 4. Different measurements to conform to guidelines or client indicated techniques 5. Custom presets of test parameters for different testing situations can be put away and later reviewed 6. Brisk survey of past estimations.
Testing Anvil - Each test hammer ought to be checked after 1000 test impacts. A testing anvil is utilized to check whether there bound test component is working accurately. If there should be an occurrence of qualities past the resistance (because of tainting by fine concrete, wear or deformities) cleaning or assessment will be required..
Ultrasonic Instrument - The pulse velocity in a material relies upon its thickness and its flexible properties, which thusly are identified with the quality and the compressive strength of the material. It is consequently conceivable to get data about the properties of parts by sonic examinations: 1. Consistency of the solid 2. Depressions, splits, absconds because of flame and ice 3. Modulus of versatility (general sign just) 4. concrete strength.