Quick answer. Heat transfer rigs measure something nobody can see, and almost every wrong student result has the same cause: heat left the apparatus by a path the equation did not include. So the question to ask of a rig is not what it demonstrates, which is printed on every brochure, but how it deals with its losses - whether each one is guarded, measured, or merely ignored. After that, three things decide whether the data is usable: the thermocouple arrangement, the heater rating and how it is controlled, and flow measurement on anything with a fluid in it. This page sets out how to specify on those grounds, and ends with an honest note about this listing, which contains roughly 22 distinct experiments spread over 57 records.
A heat transfer experiment is an energy balance. The student measures the electrical power into a heater and some temperatures, and calculates a conductivity, a coefficient or an emissivity on the assumption that the measured power went where the theory says it went. It never all does. Some leaves through the mountings, some down the thermocouple wires, some off the back of the specimen, some by radiation when the calculation assumed only convection.
There are only three honest ways to deal with a loss, and one dishonest one.
The teaching value of this equipment is in making students confront that fourth option. A rig that hides its losses produces tidy numbers and no understanding. A rig that exposes them - by giving the student the extra thermocouples and asking for a loss estimate - produces a worse-looking result and an engineer. Specify for the second.
Read this column by column rather than row by row. The third column is the one a quotation never mentions and the one that decides your data.
| Experiment | Quantity measured | Loss that dominates | How a good rig handles it |
|---|---|---|---|
| Conduction through a composite wall | Thermal resistance in series; interface resistance | Edge losses radially outward from the stack | Guard ring or generous lagging, plus a stated edge-loss bound |
| Conductivity of a metal rod | Axial conductivity from a temperature gradient | Radial loss along the whole rod | Heavy lagging, a thermocouple on the lagging surface, and a water-cooled far end whose duty is measured |
| Conductivity of an insulating powder | Conductivity of a poor conductor | Departure from the assumed spherical or cylindrical symmetry | Concentric geometry held true, and a stated fill procedure - the packing density is part of the measurement |
| Guarded hot plate on a slab | Conductivity of a slab to a recognised method | Lateral flow out of the main heater | A separately controlled guard heater with its own differential thermocouple |
| Lagged pipe | Conductivity and critical thickness of lagging | Axial loss along the pipe and out of the ends | End sections heated separately, or long enough that the central section is one-dimensional |
| Pin fin | Fin efficiency; natural against forced convection | Conduction into the base plate, and radiation at higher temperatures | A base thermocouple so base temperature is measured rather than assumed |
| Natural convection | Convective coefficient on a vertical cylinder or plate | Radiation from the hot surface, and any air movement in the room | A surroundings thermocouple so the radiation term can be subtracted, and an enclosure or shield |
| Forced convection | Coefficient against Reynolds number in a duct | Loss through the duct wall; unknown velocity profile at the test section | A lagged duct, a calibrated orifice or anemometer, and a long enough entry length |
| Emissivity comparison | Emissivity of a test plate relative to a reference | Convection from both plates, and any difference between them other than emissivity | Identical geometry and identical power, inside an enclosure, with the reference coating specified |
| Stefan-Boltzmann | The fourth-power law and the constant | Conduction into the test disc holder; non-black enclosure | A thin, low-mass disc, a hemispherical enclosure at a measured temperature |
| Heat exchanger, any type | Overall coefficient, LMTD, effectiveness | Loss to the room from pipework, and error in the flow measurement | Lagged pipework and a flow measurement whose calibration fluid and temperature are stated |
| Dropwise and filmwise condensation | The step change in coefficient between the two modes | Non-condensable gas in the shell, which destroys the result | A venting arrangement and a procedure for purging before readings |
Almost every experiment in this listing is a steady-state measurement, and the single most common procedural failure is that readings are taken when the timetable runs out rather than when the apparatus has settled. A heavily lagged pipe or a thick insulating slab has a long time constant by design - the better the lagging, the longer the wait - so the apparatus that gives the cleanest result is the one students are most likely to read too early.
The fix is to make steady state a stated criterion rather than a judgement. A department should write a tolerance into the lab sheet: for example, no measured temperature changing by more than 0.2 °C over five minutes, on at least three consecutive readings. The particular numbers belong to your apparatus and your tolerance for lab time, and should be established once by running one experiment to true equilibrium and watching the approach; the point is that a criterion exists and is written down, so that a student who is running late has to record that the criterion was not met rather than quietly taking the reading anyway.
Two practical consequences for specification:
On this equipment the sensors are not an accessory. Every number the student calculates is a difference between two temperatures, so an error in one reading propagates straight into the result, and the usual errors are mechanical rather than electrical.
A natural convection apparatus beside a door, under a ceiling fan, or in front of an air conditioner is a forced convection apparatus. The whole experiment rests on the air moving only because the surface heated it, and a draught of a fraction of a metre per second is enough to change the coefficient materially. This is worth deciding before the bench layout is fixed: the natural convection rig wants a corner away from doors and vents, and ideally a shield or enclosure. It is also the experiment where the radiation term is largest relative to the convection being measured, so the surroundings thermocouple is not optional - without it the student is attributing radiation to convection and getting a coefficient that is too high.
The teaching point is spectacular - the coefficient changes by an order of magnitude between the two modes on the same surface - and it is destroyed by air in the shell. Non-condensable gas blankets the condensing surface and collapses the coefficient, so a rig without a proper vent and a written purge procedure will show a difference far smaller than the real one, or none. Ask specifically how the shell is vented and what the start-up procedure is.
Every exchanger rig in this listing gives the student a free check on their own data, and most lab sheets do not use it. The duty calculated on the hot side and the duty calculated on the cold side are measurements of the same quantity, so they should agree within a stated tolerance.
As with any teaching laboratory, specify by ratio rather than by item. Decide the class size, the batch size at one rig, and the number of weeks in the rotation; those three numbers determine how many of each experiment you need, and the answer is almost never one of everything.
A defensible core for an undergraduate heat transfer course, in buying order:
Shared services matter more than any single rig: a stable supply, a reliable cooling-water supply at a known temperature, and a data logger used across rigs. See shared engineering lab instruments for the pumps, gauges and general instruments these rigs assume you already have.
We would rather say this than let you discover it halfway down page four. This listing holds 57 records, HTL-0001 to HTL-0057, which resolve to roughly 22 distinct experiments. Several experiments appear three times under different codes - the emissivity apparatus, the Stefan-Boltzmann apparatus, the pin fin, the forced convection unit and the dropwise/filmwise condensation unit each have three records. Two codes are used twice over (HTL-0039 and HTL-0045), HTL-0034 is missing, and a few names carry typing errors that have travelled with the data: Insulting Powder for insulating, Citical Radius for critical, Parrallel, and Stefan Boltzman. One record, HTL-0049 Heat Transfer Sheet, is not a product.
So do not order by code from this page. Tell us the experiment and the class size and we will quote the current model, confirm the specification in writing, and tell you which code it ships against. We have recommended internally that this listing be consolidated, and the duplicate records removed rather than renamed.
Within physics lab for engineering students: thermodynamics for compressors, calorimeters, steam and refrigeration-cycle rigs, hydraulics for the bench, Bernoulli and notch experiments, and analytical instruments for pH, conductivity and photometry. Momentum transfer lab equipment covers the fluid-flow side of the same transport-phenomena syllabus. The full engineering lab equipment category lists the rest.
Because heat left by a path the equation ignored, and almost always radially through the lagging rather than axially along the rod. The measured power is divided by a gradient that is steeper than the axial flow alone would produce, so the conductivity comes out low. The apparatus is usually working correctly. The fix is not a better rig but a better calculation: measure the lagging surface temperature and the far-end cooling duty, estimate the loss, and subtract it. That estimate is the most valuable thing in the experiment.
A guard heater surrounds the main heater and is held at the same temperature, so there is no sideways temperature difference and no sideways heat flow - which means all the measured power goes through the specimen. It removes the error instead of correcting for it, and it is the recognised method for insulating slabs. It is worth paying for if you need defensible conductivity values or you teach the standard method. If the aim is to show that insulation resists heat flow, an unguarded apparatus with honest loss accounting teaches that at lower cost.
Longer than the timetable allows, which is why it needs a written criterion rather than a judgement. Establish it once by running an experiment to true equilibrium and watching the approach, then write a tolerance into the lab sheet - no temperature changing by more than a stated amount over a stated interval, on several consecutive readings. Well-lagged apparatus takes longest, so your cleanest experiment is the one most likely to be read too early. Running at lower heater power shortens the wait and shrinks the losses at the same time.
In a corner away from doors, windows, ceiling fans and air-conditioner outlets, ideally with a shield or enclosure. The experiment assumes the air moves only because the surface heated it, and a slight draught turns it into a forced convection experiment with an unknown velocity. Decide this before the bench layout is fixed, and specify the surroundings thermocouple, because radiation is a large fraction of the total here and without that channel the student attributes it to convection.
Close the energy balance. The duty on the hot side and the duty on the cold side measure the same quantity, so compute both every run; a gap beyond your stated tolerance means something is wrong before any coefficient is believable. When they disagree, suspect the flow measurement rather than the exchanger - a rotameter calibrated for cold water reads wrong on hot - and check the pipework lagging. Verify one flow point against a measuring tank and a stopwatch.
Because it does. The listing has grown to 57 records covering about 22 experiments, with duplicate entries for the emissivity, Stefan-Boltzmann, pin fin, forced convection and condensation units, two codes used twice, and one missing code. We have said so on the page rather than leaving you to find it, and we have recommended consolidation. In the meantime, enquire by experiment and class size rather than by code, and we will confirm the model and specification in writing.
Ray Export is a manufacturer and exporter of laboratory, scientific and educational equipment based at Plot No. 449, HSIIDC Industrial Area, Saha, Ambala, Haryana 133104, India. Enquiries: [email protected], +91-9817319615.
Written by Palak Mall, Senior Content Writer. Technically reviewed by Arvind Kumar, Lab Equipment Specialist, 12+ years in laboratory and testing instrumentation. Last updated: 17 October 2026.
The steady-state tolerance given in this page is an illustrative criterion for a department to set and verify against its own apparatus, not a measured property of any product. Thermocouple type characteristics are given at the level of standard reference practice; the applicable temperature range, tolerance class and compensation arrangement for a specific unit will be confirmed in writing on the quotation, as will heater ratings, insulation, flow-measurement calibration conditions, services and footprint. Record counts and product-name defects described above were read from this listing on 15 October 2026 and will change as the listing is corrected. No price is quoted on this page; prices for this equipment depend on specification, instrumentation and destination and are issued as dated written quotations.
Warm Radiation Unit - Engineering Lab Training Systems ccording to the Stefan-Boltzmann law, the vitality emanated by a blackbody radiator for every second per unit region is relative to the fourth energy of the supreme temperature. Warmth exchange through radiation happens in type of electromagnetic waves for the most part in the infrared area. Radiation transmitted by a body is an outcome of warm disturbance of its forming particles. This lab gadget is worked to figure the aggregate vitality transmitted per unit surface territory of the dark body by the Stefan-Boltzmann law..
Two Pipe Heat Exchanger - Engineering Lab Training Systems The twofold tube warm exchanger is the easiest type of warmth exchanger, comprising of two concentric (coaxial) tubes, one conveying the hot liquid and the other conveying the icy liquid. Warmth is exchanged from the hot to the cool liquid without blending of the two streams. In the present lab setup, four fragments of twofold pipe warm exchangers are masterminded in arrangement, where temperatures are estimated at the channel and outlet of each section. A progression of sidestep tubes are utilized to permit the task of one, two, three, or four sections, in parallel or counter stream setups, as required by the particular lab works out. A progression of valves control the stream rates and the course of move through a control program..
Free And Forced Convection Unit-Engineering Lab Training Systems Thisexperimental unit is intended for nothing and constrained convection. Air is nourished through a conduit and warmed at radiator components with various formed surfaces.Three exchangeable warmer components are given. A fan can be exchanged on for probes constrained convection. The temperature and air speed are estimated utilizing sensors..
Warmth Conduction In Gases And Liquids - Engineering Lab Training Systems General Bench top unit for the assurance of warm conductivity of fluid and gasses, and finish with a support for the control and show of temperature and warmth input..
Dropwise Filmwise Condensation Apparatus Condensation is a stage change warm move process happening in numerous mechanical applications, for example, in steam control plants, refrigeration plants and so on. Consequently this is one of the imperative warmth move process introduce in mechanical and substance designing application. The buildup of Vapor on a surface is of two kinds: 1. Drop astute Condensation. 2. Film savvy Condensation. This set up is intended for discovering heat exchange co-proficient of two kinds of buildup and for representation of these procedures. It comprises of a vertical edge. Buildup tubes are fitted inside conservative glass barrel. Steam generator is fitted at the rear of the chamber. Steam comes specifically from generator to chamber. Two valves are fitted to control stream rate of water in singular tubes. Computerized Temperature Indicator screens temperatures. Weight check and Rotameter can watch steam weight and cool, water stream rates separately. A Digital temperature Controller is accommodated controlling the temperature of Steam. Water level pointer is given to defend the warmer. Condensate is estimated by estimating barrel. Investigations To think about the Film shrewd and Drop astute buildup of steam on a vertical surface Visualization of buildup process in drop insightful and additionally film savvy Utilities Required Water supply 5 lit/min (approx.) Electricity Supply: 1Phase. 220 V AC. 1.5 kW Table for set-up help Heat Transfer lab hardware manufacturers,exporters and Heat Transfer Lab providers
Stick Fin Apparatus The Pin Fin Apparatus is intended to decide warm exchange from an expanded surface (Pin). It comprises of a metallic stick blade which is warmed toward one side by electric warmer. The warmth is led along the length & is streams out from the stick surface. Five thermocouples are implanted along the length of the stick to gauge the temperature dispersion, one more thermocouple is set inside the suction conduit to quantify the air temperature. A blower gives constrained convection of air by blowing the air at high speed. Stream rate of blower is estimated by opening plate & U Tube Manometer. The control board comprises of a dimmers detail to control the voltage provided to the warming component. A Voltmeter, an ammeter and a computerized temperature marker are additionally given. Particulars: Pin: MOC Aluminum, Pin width, 15 mm, Length 20mm Electric Heater: Capacity 200 W, Supply 230 V AC Air Blower: Centrifugal Blower, 230 V AC Orifice Plate: For Air Flow estimation Manometer: U tube manometer, 0-300 mm WC Dimmer: Range 0-250 V AC, 2 A Digital Voltmeter: Range 0-500 V AC Digital Ammeter: Range 0-2 An AC Temperature Sensors: Type ‘K’ Thermocouples – 6Nos Temperature Indicator: 8 Channel Indicator with Selector Switch, Range 0-199.9 ° Frame: Made of M.S. Square Tubes & Sheets, Welded & Powder covered Experimental Capabilities: To examine marvel of Heat Transfer from expanded surface (Pin) To think about the temperature appropriation through a stick blade tentatively & hypothetically .
Constrained Convection Apparatus The Forced Convection Apparatus is intended to decide warm move rate in constrained convection. It comprises of even copper tube warmed by an electric radiator. Number of thermocouples are fitted on the surface of the tube along its length. The test pipe is housed inside a steel pipe. A blower gives constrained convection of air by blowing the air at high speed to cause turbulent stream. Stream rate of blower is estimated by hole plate & U Tube Manometer.The control board comprises of a dimmers detail to control the voltage provided to the warming component. A Voltmeter, an ammeter and a computerized temperature pointer are likewise given. Particulars: Test Pipe: Copper Pipe, Diameter 30 mm, Length of Pipe 500 mm Electric Heater: Capacity 300 W, Supply 230 V AC Air Blower: Centrifugal Blower, 230 V AC Orifice Plate: For Air Flow estimation Manometer: U tube manometer, 0-300 mm WC Dimmer: Range 0-250 V AC, 2 A Digital Voltmeter: Range 0-500 V AC Digital Ammeter: Range 0-2 An AC Temperature Sensors: Type ‘K’ Thermocouples – 12 Nos Temperature Indicator: 12 Channel Indicator with Selector Switch, Range 0-199.9 ° Frame: Made of M.S. Square Tubes & Sheets, Welded & Powder covered Experimental Capabilities: To think about wonder of Heat Transfer in Forced Convection To decide the Heat Transfer Coefficient (h) in Forced Convection Comparison of trial comes about with hypothetical qualities .