Quick answer. Heat transfer rigs measure something nobody can see, and almost every wrong student result has the same cause: heat left the apparatus by a path the equation did not include. So the question to ask of a rig is not what it demonstrates, which is printed on every brochure, but how it deals with its losses - whether each one is guarded, measured, or merely ignored. After that, three things decide whether the data is usable: the thermocouple arrangement, the heater rating and how it is controlled, and flow measurement on anything with a fluid in it. This page sets out how to specify on those grounds, and ends with an honest note about this listing, which contains roughly 22 distinct experiments spread over 57 records.

The framework: account for every watt that leaves

A heat transfer experiment is an energy balance. The student measures the electrical power into a heater and some temperatures, and calculates a conductivity, a coefficient or an emissivity on the assumption that the measured power went where the theory says it went. It never all does. Some leaves through the mountings, some down the thermocouple wires, some off the back of the specimen, some by radiation when the calculation assumed only convection.

There are only three honest ways to deal with a loss, and one dishonest one.

  • Guard it. Arrange the apparatus so the loss path carries no heat, because there is no temperature difference across it. This is what the word guarded means on a guarded hot plate: a surrounding guard heater is held at the same temperature as the main heater, so no heat flows sideways and all the measured power goes through the specimen. It is the most expensive solution and the only one that removes the error rather than correcting for it.
  • Measure it. Instrument the loss path and subtract. A thermocouple at the far end of a conductivity rod tells you what is leaving the end; a thermocouple on the outside of the lagging tells you what is leaving radially.
  • Bound it. Show by calculation that the loss is small compared with the flow being measured, and quote the bound. This is legitimate and it is what good lab manuals do, but it only works if the bound is actually computed rather than asserted.
  • Ignore it. The default. It is the reason a student measuring the thermal conductivity of brass reports a value well below the book figure, concludes the apparatus is broken, and learns nothing - when in fact the apparatus behaved exactly as physics requires and the equation was the thing that was wrong.

The teaching value of this equipment is in making students confront that fourth option. A rig that hides its losses produces tidy numbers and no understanding. A rig that exposes them - by giving the student the extra thermocouples and asking for a loss estimate - produces a worse-looking result and an engineer. Specify for the second.

What each experiment actually measures, and what it loses

Read this column by column rather than row by row. The third column is the one a quotation never mentions and the one that decides your data.

ExperimentQuantity measuredLoss that dominatesHow a good rig handles it
Conduction through a composite wallThermal resistance in series; interface resistanceEdge losses radially outward from the stackGuard ring or generous lagging, plus a stated edge-loss bound
Conductivity of a metal rodAxial conductivity from a temperature gradientRadial loss along the whole rodHeavy lagging, a thermocouple on the lagging surface, and a water-cooled far end whose duty is measured
Conductivity of an insulating powderConductivity of a poor conductorDeparture from the assumed spherical or cylindrical symmetryConcentric geometry held true, and a stated fill procedure - the packing density is part of the measurement
Guarded hot plate on a slabConductivity of a slab to a recognised methodLateral flow out of the main heaterA separately controlled guard heater with its own differential thermocouple
Lagged pipeConductivity and critical thickness of laggingAxial loss along the pipe and out of the endsEnd sections heated separately, or long enough that the central section is one-dimensional
Pin finFin efficiency; natural against forced convectionConduction into the base plate, and radiation at higher temperaturesA base thermocouple so base temperature is measured rather than assumed
Natural convectionConvective coefficient on a vertical cylinder or plateRadiation from the hot surface, and any air movement in the roomA surroundings thermocouple so the radiation term can be subtracted, and an enclosure or shield
Forced convectionCoefficient against Reynolds number in a ductLoss through the duct wall; unknown velocity profile at the test sectionA lagged duct, a calibrated orifice or anemometer, and a long enough entry length
Emissivity comparisonEmissivity of a test plate relative to a referenceConvection from both plates, and any difference between them other than emissivityIdentical geometry and identical power, inside an enclosure, with the reference coating specified
Stefan-BoltzmannThe fourth-power law and the constantConduction into the test disc holder; non-black enclosureA thin, low-mass disc, a hemispherical enclosure at a measured temperature
Heat exchanger, any typeOverall coefficient, LMTD, effectivenessLoss to the room from pipework, and error in the flow measurementLagged pipework and a flow measurement whose calibration fluid and temperature are stated
Dropwise and filmwise condensationThe step change in coefficient between the two modesNon-condensable gas in the shell, which destroys the resultA venting arrangement and a procedure for purging before readings

Steady state is a decision, not an event

Almost every experiment in this listing is a steady-state measurement, and the single most common procedural failure is that readings are taken when the timetable runs out rather than when the apparatus has settled. A heavily lagged pipe or a thick insulating slab has a long time constant by design - the better the lagging, the longer the wait - so the apparatus that gives the cleanest result is the one students are most likely to read too early.

The fix is to make steady state a stated criterion rather than a judgement. A department should write a tolerance into the lab sheet: for example, no measured temperature changing by more than 0.2 °C over five minutes, on at least three consecutive readings. The particular numbers belong to your apparatus and your tolerance for lab time, and should be established once by running one experiment to true equilibrium and watching the approach; the point is that a criterion exists and is written down, so that a student who is running late has to record that the criterion was not met rather than quietly taking the reading anyway.

Two practical consequences for specification:

  • Ask for a data logger, or at least a timer and a tabulated reading sheet, on the slow experiments. A rig with a multi-channel logger lets a class start an experiment, do something else, and come back - which is the only way a two-hour time constant fits in a one-hour slot.
  • Prefer a heater with a stable, measurable input over a powerful one. Steady state is reached sooner at low power, the losses are smaller in absolute terms, and a dimmerstat plus a true-RMS wattmeter reading is worth more than a bigger element. Ask what the power measurement is: a voltmeter and ammeter on a phase-angle-controlled supply do not multiply to give the real power.

The thermocouple arrangement is the measurement

On this equipment the sensors are not an accessory. Every number the student calculates is a difference between two temperatures, so an error in one reading propagates straight into the result, and the usual errors are mechanical rather than electrical.

  • Embedded, not resting. A junction in a drilled hole with thermal paste or a press fit reads the metal. A bead taped to a surface reads something between the surface and the air, and on a hot surface in still air that can be far from the surface temperature. Ask how each junction is mounted, specimen by specimen.
  • Type, and why. Type K (nickel-chromium against nickel-aluminium) covers the whole range of this equipment comfortably and is the sensible default. Type J (iron against constantan) is cheaper and adequate at lower temperatures but the iron leg corrodes. Mixing types on one selector switch without matching the compensation is a classic source of a constant offset on some channels and not others.
  • Count the channels you actually need, including the ones for the losses: surroundings temperature for anything radiating, lagging-surface temperature for anything lagged, inlet and outlet on both sides of any exchanger, and the far end of any conductivity rod. A rig sold with six channels for an experiment that needs eight will be used with six.
  • Calibration is a purchasable experiment. A thermocouple calibration rig turns a maintenance task into a lab exercise, and it is the honest answer to a student who asks how we know the thermocouple is right. If your department is buying a set, one calibration rig serves the whole set.

Two experiments that are nearly always sited wrongly

Natural convection

A natural convection apparatus beside a door, under a ceiling fan, or in front of an air conditioner is a forced convection apparatus. The whole experiment rests on the air moving only because the surface heated it, and a draught of a fraction of a metre per second is enough to change the coefficient materially. This is worth deciding before the bench layout is fixed: the natural convection rig wants a corner away from doors and vents, and ideally a shield or enclosure. It is also the experiment where the radiation term is largest relative to the convection being measured, so the surroundings thermocouple is not optional - without it the student is attributing radiation to convection and getting a coefficient that is too high.

Dropwise and filmwise condensation

The teaching point is spectacular - the coefficient changes by an order of magnitude between the two modes on the same surface - and it is destroyed by air in the shell. Non-condensable gas blankets the condensing surface and collapses the coefficient, so a rig without a proper vent and a written purge procedure will show a difference far smaller than the real one, or none. Ask specifically how the shell is vented and what the start-up procedure is.

Heat exchangers: the energy balance is the built-in self-check

Every exchanger rig in this listing gives the student a free check on their own data, and most lab sheets do not use it. The duty calculated on the hot side and the duty calculated on the cold side are measurements of the same quantity, so they should agree within a stated tolerance.

  • If they do not agree, suspect the flow measurement first. Temperatures are easy and flows are hard. A rotameter is calibrated for a stated fluid at a stated temperature and density; running hot water through a float calibrated for cold water gives a reading that is wrong in a direction most students will not question. Ask what the calibration conditions are, and prefer a measuring tank and stopwatch as the reference method for at least one point.
  • Lag the pipework, not just the exchanger. On a small rig the pipe runs can have comparable surface area to the exchanger, and the loss from them appears as a failure of the energy balance that students will blame on the exchanger.
  • Parallel against counter flow is the experiment, not the exchanger type. A rig with valved reversal lets one unit teach both and makes the LMTD comparison immediate; two separate units teach the same thing at twice the cost and with the configurations never compared on the same day under the same conditions. If you buy one exchanger for a teaching lab, buy the one that reverses.
  • Effectiveness-NTU is the better framing for a design course, and it needs the same measurements, so a rig adequate for LMTD is adequate for both - but check that the manual actually covers it, because many do not.

Specifying a set for a class

As with any teaching laboratory, specify by ratio rather than by item. Decide the class size, the batch size at one rig, and the number of weeks in the rotation; those three numbers determine how many of each experiment you need, and the answer is almost never one of everything.

A defensible core for an undergraduate heat transfer course, in buying order:

  1. Conduction: composite wall or insulating slab, and a metal rod. These establish the series-resistance idea and the loss problem.
  2. Convection: natural and forced, and the pin fin which bridges the two. Duplicate these first when the class grows, because they are the most used.
  3. Radiation: emissivity comparison and Stefan-Boltzmann. One of each is usually enough.
  4. Application: one reversible exchanger, plus a shell-and-tube or plate unit if the course covers industrial equipment.
  5. Everything else - evaporators, heat pipe, unsteady state, critical radius, agitated vessel - is a second phase, bought against specific syllabus requirements rather than to fill a lab.

Shared services matter more than any single rig: a stable supply, a reliable cooling-water supply at a known temperature, and a data logger used across rigs. See shared engineering lab instruments for the pumps, gauges and general instruments these rigs assume you already have.

What a quotation must state

  • Heater rating and how power is measured - element wattage, the controller type, and whether the instrument reads true power.
  • Thermocouple type, count, and mounting method for each junction, including any channel provided for losses or surroundings.
  • The guard arrangement, where the method depends on one, and whether the guard has independent control.
  • Insulation material and thickness, and whether the apparatus provides a means to measure the loss through it.
  • Flow measurement device, its range, and its calibration fluid and temperature.
  • Services required - supply voltage and phase, cooling water flow and pressure, compressed air, drainage.
  • Bench footprint and working height, and whether the unit is free-standing or needs a bench.
  • What is included - specimens, spare elements, the manual, and whether sample calculations with real data are provided.

About this listing - read before you order by code

We would rather say this than let you discover it halfway down page four. This listing holds 57 records, HTL-0001 to HTL-0057, which resolve to roughly 22 distinct experiments. Several experiments appear three times under different codes - the emissivity apparatus, the Stefan-Boltzmann apparatus, the pin fin, the forced convection unit and the dropwise/filmwise condensation unit each have three records. Two codes are used twice over (HTL-0039 and HTL-0045), HTL-0034 is missing, and a few names carry typing errors that have travelled with the data: Insulting Powder for insulating, Citical Radius for critical, Parrallel, and Stefan Boltzman. One record, HTL-0049 Heat Transfer Sheet, is not a product.

So do not order by code from this page. Tell us the experiment and the class size and we will quote the current model, confirm the specification in writing, and tell you which code it ships against. We have recommended internally that this listing be consolidated, and the duplicate records removed rather than renamed.

Related ranges

Within physics lab for engineering students: thermodynamics for compressors, calorimeters, steam and refrigeration-cycle rigs, hydraulics for the bench, Bernoulli and notch experiments, and analytical instruments for pH, conductivity and photometry. Momentum transfer lab equipment covers the fluid-flow side of the same transport-phenomena syllabus. The full engineering lab equipment category lists the rest.

Frequently asked questions

Why does a student measuring thermal conductivity get a value below the book figure?

Because heat left by a path the equation ignored, and almost always radially through the lagging rather than axially along the rod. The measured power is divided by a gradient that is steeper than the axial flow alone would produce, so the conductivity comes out low. The apparatus is usually working correctly. The fix is not a better rig but a better calculation: measure the lagging surface temperature and the far-end cooling duty, estimate the loss, and subtract it. That estimate is the most valuable thing in the experiment.

What does "guarded" mean on a hot plate apparatus, and is it worth paying for?

A guard heater surrounds the main heater and is held at the same temperature, so there is no sideways temperature difference and no sideways heat flow - which means all the measured power goes through the specimen. It removes the error instead of correcting for it, and it is the recognised method for insulating slabs. It is worth paying for if you need defensible conductivity values or you teach the standard method. If the aim is to show that insulation resists heat flow, an unguarded apparatus with honest loss accounting teaches that at lower cost.

How long should an experiment take to reach steady state?

Longer than the timetable allows, which is why it needs a written criterion rather than a judgement. Establish it once by running an experiment to true equilibrium and watching the approach, then write a tolerance into the lab sheet - no temperature changing by more than a stated amount over a stated interval, on several consecutive readings. Well-lagged apparatus takes longest, so your cleanest experiment is the one most likely to be read too early. Running at lower heater power shortens the wait and shrinks the losses at the same time.

Where should the natural convection apparatus be installed?

In a corner away from doors, windows, ceiling fans and air-conditioner outlets, ideally with a shield or enclosure. The experiment assumes the air moves only because the surface heated it, and a slight draught turns it into a forced convection experiment with an unknown velocity. Decide this before the bench layout is fixed, and specify the surroundings thermocouple, because radiation is a large fraction of the total here and without that channel the student attributes it to convection.

How do we know a heat exchanger rig is giving good data?

Close the energy balance. The duty on the hot side and the duty on the cold side measure the same quantity, so compute both every run; a gap beyond your stated tolerance means something is wrong before any coefficient is believable. When they disagree, suspect the flow measurement rather than the exchanger - a rotameter calibrated for cold water reads wrong on hot - and check the pipework lagging. Verify one flow point against a measuring tank and a stopwatch.

Why does this category list several records that look like the same experiment?

Because it does. The listing has grown to 57 records covering about 22 experiments, with duplicate entries for the emissivity, Stefan-Boltzmann, pin fin, forced convection and condensation units, two codes used twice, and one missing code. We have said so on the page rather than leaving you to find it, and we have recommended consolidation. In the meantime, enquire by experiment and class size rather than by code, and we will confirm the model and specification in writing.

Key takeaways

  • Account for every watt that leaves. Each loss must be guarded, measured, or bounded by calculation - and the fourth option, ignoring it, is what produces wrong student results on working apparatus.
  • The thermocouple arrangement is the measurement. Embedded junctions, one consistent type, and enough channels to instrument the losses as well as the specimen.
  • Make steady state a written criterion, not a judgement, and prefer a stable low heater power with true power measurement over a bigger element.
  • Site the natural convection rig away from every draught, and buy the surroundings thermocouple with it - radiation is a large share of the total.
  • Close the energy balance on every exchanger run. When the two sides disagree, the flow measurement is the usual culprit; buy the exchanger that reverses between parallel and counter flow.
  • Specify by ratio - class size, batch size, weeks in the rotation - and buy conduction, convection and radiation cores before the specialist units.

About this page

Ray Export is a manufacturer and exporter of laboratory, scientific and educational equipment based at Plot No. 449, HSIIDC Industrial Area, Saha, Ambala, Haryana 133104, India. Enquiries: [email protected], +91-9817319615.

Written by Palak Mall, Senior Content Writer. Technically reviewed by Arvind Kumar, Lab Equipment Specialist, 12+ years in laboratory and testing instrumentation. Last updated: 17 October 2026.

The steady-state tolerance given in this page is an illustrative criterion for a department to set and verify against its own apparatus, not a measured property of any product. Thermocouple type characteristics are given at the level of standard reference practice; the applicable temperature range, tolerance class and compensation arrangement for a specific unit will be confirmed in writing on the quotation, as will heater ratings, insulation, flow-measurement calibration conditions, services and footprint. Record counts and product-name defects described above were read from this listing on 15 October 2026 and will change as the listing is corrected. No price is quoted on this page; prices for this equipment depend on specification, instrumentation and destination and are issued as dated written quotations.

Stefan Boltzmann Apparatus

HTL-0011

The mechanical assembly is intended to decide, the Stefan Boltzmann consistent. The device comprises of a side of the equator settled to a Bakelite plate, the external surface of which frames the coat to warm it. High temp water to warm the half of the globe is gotten shape a heated water tank, which is settled over the side of the equator. The copper test plate is presented at the focal point of half of the globe. The temperatures of side of the equator and test plate are estimated with the assistance of temperature sensors. 


Tests 


Assurance of Stefan Boltzmann consistent and ponder the impact of side of the equator temperature on it 


Utilities Required 


Power Supply: 1Phase, 220 VAC, 4 Amp. 


Table for set-up help

Dropwise / Filmwise Condensation Apparatus

HTL-0012

Buildup is a stage change warm move process happening in numerous mechanical applications, for example, in steam control plants, refrigeration plants and so on. Therefore this is one of the imperative warmth move process exhibit in mechanical and synthetic designing application. The buildup of Vapor on a surface is of two sorts: 1. Drop insightful Condensation. 2. Film savvy Condensation. This set up is intended for discovering heat exchange co-proficient of two kinds of buildup and for perception of these procedures. It comprises of a vertical casing. Buildup tubes are fitted inside conservative glass chamber. Steam generator is fitted at the posterior of the chamber. Steam comes specifically from generator to barrel. Two valves are fitted to control stream rate of water in singular tubes. Advanced Temperature Indicator screens temperatures. Weight measure and Rotameter can watch steam weight and cool, water stream rates separately. A Digital temperature Controller is accommodated controlling the temperature of Steam. Water level marker is given to protect the radiator. Condensate is estimated by estimating barrel. 


Investigations 


To ponder the Film savvy and Drop insightful buildup of steam on a vertical surface 


Perception of buildup process in drop savvy and additionally film shrewd 


Utilities Required 


Water supply 5 lit/min (approx.) 


Power Supply: 1Phase. 220 V AC. 1.5 kW 


Table for set-up help

Shell & Tube Heat Exchanger - Engineering Lab Equipment

HTL-0013

Shell and Tube Heat Exchanger are well known in enterprises since they possess less space and offer sensible temperature drop. The device comprises of created SS shell, inside which copper tubes with confuses on external side are fitted. This is two-pass warm exchanger with the goal that high temp water goes to one end of shell through the tubes and comes back to another end through outstanding tubes. The icy water is conceded at the one end of shell, which ignores the high temp water tubes. Valves are given to control the stream rates of hot and frosty water. Stream rates of hot and cool water are estimated utilizing Rotameters. An attractive drive pump is utilized to circle the boiling water from a re-cycled type water tank, which is fitted with radiators and Digital Temperature Controller. 


Analyses 


The principle question of the test setup is to consider the accompanying at different stream rates: 


LMTD. . 


Warmth exchange rate 


General Heat Transfer Co-productive 


Utilities Required 


Water supply 20 lit/min (approx.) and deplete 


Power Supply: I Phase, 220 V AC, 4 kW 


Floor zone of 1.5m x 0.75 m

Heat Transfer In Agitated Vessel

HTL-0014

In most mechanical procedures including reactors, warm is to be added or separated to control the procedure. The expansion and evacuation of warmth is finished by passing steam in coat fitted to the outside of the vessel or passing chilly. Water in helical loop inside the vessel. For compelling warmth exchange and even dissemination of warmth, the fluid inside is ceaselessly disturbed. The present set-up offers us. A near investigation of Jacket and Helical curl, which can be worked at the same time. It is a stainless steel jacketed vessel inside which, a helical loop of copper is fitted. A variable speed stirrer is fitted in vessel. The framework is such composed that either steam or water is permitted to enter inside anybody of the coat and helical loop by a course of action of control valves. Variety in temperature of inside water is estimated and is noted. Set-up is fitted with steam traps for condensate accumulation. 


Stream rate of water can be controlled and estimated utilizing valve and Rotameter. Temperature of gulf and outlet of both hot and cool liquids are estimated by temperature sensors. 


Tests 


To decide the general warmth exchange co-effective for different level of unsettling 


To make a relative investigation of Heat Transfer through Jacket and Coil in an unsettled vessel 


Utilities Required 


Water supply 20 lit/min (approx.) and Drain. 


Power Supply: 1 Phase, 220 V AC, 5.kW. 


Floor territory of 1.5m x 1.5m

Unsteady State Heat Transfer Unit

HTL-0015

The mechanical assembly comprises of a little test chamber. The barrel is warmed by a consistent temperature water shower, till relentless state is come to. Amid warming, temperature of the barrel is capacity of time and thus, warming of chamber is precarious state warm exchange. The temperature of chamber is estimated with the assistance of temperature sensor embedded in the middle. The boiling water shower is furnished with a radiator and is controlled by computerized temperature controller. An instigator is likewise given to keep up the consistent temperature. 


Examinations 


To compute the Biot number, Fourier number and to figure the warmth exchange coefficient 


To plot Heisler Chart for the given sa1l1pleof round and hollow strong 


Utilities Required 


Power Supply: 1 Phase, 220 V AC, I kW. 


Seat zone of 1.5m x 1.5m

Multi Effect Evaporator

HTL-0016

Dissipation manages the centralization of a non-unpredictable Solute from an answer by the evacuation of required measure of unstable dissolvable. Normally the dissolvable is water. By vaporizing a piece of the dissolvable, helpful item i.e. the concentrated arrangement or thick alcohol is delivered and the Vapor is disposed of. Lung tube evaporators are normally utilized for the convergence of frothy fluids. The set-up is comprises of two evaporators fitted in arrangement. Each is comprised of Stainless Steel tubes encompassed by a Stainless Steel coat and fitted with aggregator. Weaken arrangement is encourage to the principal evaporator. Steam from. a steam generator is provided to evaporator to think the weaken encourage answer for a coveted level. The coat is fitted with a steam trap and the condensate is gathered toward the finish of trap. The vapors of unstable dissolvable delivered in first evaporator are provided to the second evaporator. The vapors of unpredictable dissolvable are consolidated in a shell and tube compose condenser and the adjust non-unstable solute gathered in the gatherer is reused through the evaporator. 


Trials 


To focus an answer and to assess the accompanying at unfaltering state: 


Material and Heat Balance 


Economy and the limit of the evaporator for various encouraging game plan 


General warmth exchange co-effective. 


Utilities Required 


Water supply 10 lit/min (approx.) and Drain 


Power Supply: 1 Phase, 220 V AC, 3-6 kW 


Required Chemicals

Citical Radius of Insulating Material

HTL-0017

We are engaged with assembling and supply of Critical Radius of Insulation material. With this mechanical assembly, the truth of the matter is contemplated that at specific range of protection over a barrel, loss of warmth is most extreme. The unit involves a G.I. pipe covered with mortar of Paris protection on outside surface. The unit has four funnels with a radiator inside having input control. Contribution to the warmers is figured by shared voltmeter and ammeter. By putting control over the contribution to radiators, it is demonstrated that basic span of protection is proportion of warm conductivity and outside warmth exchange co-effective.

Specific Heat of Air Apparatus

HTL-0018

1. The particular warmth of air mechanical assembly has a warming component for the air to be blown upon. 


2. The distinction in temperature of air is marker with the assistance of temperature pointer. 


3. The stream of air is recorded in the opening meter. 


4. The contribution to the radiator is estimated by voltmeter and ammeter. In this manner the particular warmth of air can be discovered.

Critical Heat Flux Apparatus

HTL-0019

Basic Heat Flux Apparatus: 


Basic Heat Flux Apparatus functions as a definite device to introduce the strategy for pool bubbling. Our motion contraption has a test wire encased by water of relentless temperature. The surface temperature of the test wire is warmed by going current through it. Because of the temperature diverse temperatures in the surface and air, the distinction in framework are seen till it wears out. 


Determinations: 


1. Test wire - 35 check or 40 measures. Length = 10cm. 


2. Nichrome warmer 1 kw limit - 1 nos. 


3. Glass trough to hold the water 


4. Table light to watch the test wire. 


5. Voltmeter and ammeter to quantify the information sent to the test wire. 


6. Dimmerstat to change the voltage of the device

Concentric Tube Heat Exchanger

HTL-0020

(A)Plain Tube (Water to Water): 


1. The device contain a couple of concentric tubes. Inward tube comprises of copper tube through which boiling water streams. 


2. While the external tube comprises of GI pipe through which icy water streams. The course of action empowers the parallel and counter stream of the frosty water. 


3. The channel and outlet temperatures and the stream estimation empower to discover the warmth traded LMTD, and viability of the hardware. 


(B) Finned Tube (Water to Air) 


1. The device includes of two concentric tubes. Internal tube comprises of copper tube through which boiling water streams. 


2. It is finned at its external measurement to upsurge the warmth exchange surface. While the external tube comprises of GI pipe through which chilly wind currents. 


3. The association empowers the parallel and counter stream of the air. The gulf and outlet temperatures and the stream estimation empower to discover the warmth traded LMTD, and effectiveness of the hardware.

Supply of Heat Transfer Lab Equipment from India to Finland, France, Tanzania, Thailand, Sri Lanka, Congo, Costa Rica, Ivory Coast, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Djibouti, Dominica, Afghanistan, Albania, Algeria, Andorra, Angola, Antigua and Barbuda, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Bangladesh, Barbados, United Arab Emirates (UAE, Dubai), United Kingdom (London), United States, Uruguay, Uzbekistan, Vanuatu, Belarus, Belgium, Belize, Benin, Bhutan, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei, Bulgaria, Burkina Faso, Burma Myanmar, East Timor, Ecuador, Egypt, El Salvador, Equatorial Guinea, Eritrea, Estonia, Ethiopia (Addis Ababa), Fiji, Burundi, Cambodia, Cameroon, Canada, Cape Verde, Central African Republic, Chad, Chile, Colombia. Math Teaching Kit suppliers in Luxembourg, Macedonia, Madagascar, Malawi (Lilongwe), Malaysia (Kuala Lumpur), Maldives, Mali, Malta, Marshall Islands, Mauritania, Mexico, Micronesia, Moldova, Monaco, Mongolia, Montenegro, Morocco, Mozambique, Namibia, Comoros, Congo. Math Teaching Kit manufacturers in Dominican Republic, Gabon, Gambia, Georgia, Germany, Ghana, Greece, Grenada, Guatemala, Guinea, Guinea-Bissau, Guyana, Haiti, Honduras, Hungary, Iceland, Indonesia, Iran, Iraq, Niger, Nigeria (Abuja), Norway, Oman, Palau, Panama, Papua New Guinea, Paraguay, Peru, Philippines (Manila), Poland, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya (Nairobi), Kiribati, Korea, Kuwait, Kyrgyzstan, Laos, Latvia, Lebanon, Lesotho, Liberia, Liechtenstein, Lithuania, Mauritius, Nauru, Nepal, Netherlands, New Zealand, Nicaragua, Portugal, Qatar, Romania, Russia, Rwanda (Kigali). Math Teaching Kit exportets to Saint Kitts and Nevis, Saint Lucia, Saint Vincent and the Grenadines, Samoa, San Marino, Sao Tome and Principe, Saudi Arabia, Senegal, Serbia, Seychelles, Sierra Leone, Singapore, Slovakia, Slovenia, Solomon Islands, Somalia, South Africa, South Sudan, Spain, Sudan, Suriname, Swaziland, Sweden, Switzerland, Syria, Tajikistan, Togo, Tonga, Trinidad and Tobago, Tunisia, Turkey, Turkmenistan, Tuvalu, Uganda (Kampala), Ukraine, Venezuela, Vietnam, Yemen , Zambia (Lusaka), Zimbabwe