Quick answer. A microwave test bench is not one instrument. It is a line of waveguide components bolted together - a source, a length of guide with the measuring sections in it, a load, and a meter - and it only works if every component is for the same waveguide band with the same flange. Ray Export supplies six benches here, MTB-0001 to MTB-0006, differing mainly in the source (reflex klystron or Gunn diode) and in what the bench is set up to measure (standing wave ratio, guide wavelength, dielectric constant, E-plane power division, horn radiation). The two decisions that matter are the source and the band, and the hazard that actually hurts people in a microwave teaching laboratory is not the microwave energy - it is the several hundred volts in a klystron supply.
Every bench in this range, and every microwave experiment, is these four blocks. Specify them as four decisions and a quotation becomes checkable.
The reflex klystron is a vacuum device. It needs a power supply providing a beam voltage of several hundred volts and an adjustable repeller voltage, and the relationship between repeller voltage and output is itself a standard experiment - the mode characteristic. It is the traditional teaching source and the one most syllabuses name.
The Gunn diode is a solid-state source running from a low voltage supply, usually with a PIN modulator to switch it. It is simpler, more robust, and markedly safer in a student laboratory, and it has its own experiment - the current-voltage characteristic and the threshold at which oscillation begins.
Our advice, which reduces the sale rather than increasing it: if your syllabus does not specifically require klystron mode characteristics, a Gunn bench is the better first purchase for a teaching laboratory. There is no high-voltage supply to maintain, no vacuum tube to replace, and nothing in the rest of the bench changes. If the syllabus does name the klystron, buy the klystron bench and treat its power supply as the laboratory's principal electrical hazard.
Waveguide is band-specific. A component for one band will not pass the frequencies of another and its flange will usually not even bolt to a component from a different band. This is the single most common procurement error in microwave teaching equipment: components bought from two sources, in good faith, that cannot be assembled.
So when you enquire, and when you compare quotations: state the waveguide band and the flange standard, and ask us to confirm both in writing for every item. If you already own a bench and are extending it, tell us what you have - we would rather match it than send you something that does not bolt on.
The measuring sections in the guide, and what each is for:
Half of microwave measurement is deciding what the guide is terminated with. A matched load absorbs everything and gives the reference condition; a movable short gives a full reflection whose position you control, which is how guide wavelength is measured; a variable or sliding load gives intermediate reflections for practice.
The point buyers miss: the quality of the matched load sets the floor on every standing wave ratio you can measure. If the load itself has a standing wave ratio of 1.05, nothing on that bench will ever read better than about 1.05, and a student who measures a well-matched component will conclude it is mismatched. Ask for the load's own figure, and treat it as a reference item - kept clean, capped, and not dropped.
This is the part of a microwave bench that confuses almost every student and quite a few technicians, and it is worth putting on the laboratory wall.
A standing wave ratio meter is not a general voltmeter. It is a narrow-band amplifier tuned to a fixed low audio frequency, conventionally 1 kHz, with a scale calibrated in standing wave ratio. For it to read anything at all, the microwave source must be square-wave modulated at that same frequency - from the klystron supply's internal modulator, or from the PIN modulator on a Gunn bench. A bench that reads nothing is usually a bench whose modulation is off or set to the wrong frequency, not a bench that is broken.
And the limitation that produces wrong answers rather than no answer: the meter scale assumes the crystal detector is operating in its square-law region, which it only does at low power. Run the bench at a high level and the detector leaves that region, the scale no longer corresponds to standing wave ratio, and the readings are wrong while looking perfectly plausible. The discipline is simple - keep the level low, use the attenuator rather than the source to set it, and if a reading looks surprisingly good or surprisingly bad, reduce the level and take it again.
One piece of physics causes more wrong answers in this subject than anything else, so we state it here rather than leaving it to the manual. Inside a waveguide the wavelength is longer than it would be in free space at the same frequency, and it depends on the guide dimensions through the cutoff wavelength. A student who measures the distance between standing-wave minima, doubles it, and treats the result as the free-space wavelength will calculate a frequency that is simply wrong.
The correct route is to measure the guide wavelength from the minima, take the cutoff wavelength from the guide dimensions, and combine the two. It is a short calculation, it is in every textbook, and it is skipped constantly. A bench that comes with a worked example of it is worth more than one that does not - ask whether the experiment manual includes it.
One administrative point to settle before purchase: a microwave bench radiates, and the band it operates in may be regulated where you are. Confirm with your national spectrum authority whether operating this equipment, even inside a laboratory, requires notification or authorisation. We will tell you the band in writing so the question can be answered.
Related ranges: microwave lab trainers for the klystron and Gunn bench items listed separately, antenna and satellite trainers, radar trainers, fibre optics trainers, communication and RF trainers and advanced communication trainers. The full training kits and electronics lab trainer kits categories list the rest.
If the syllabus names klystron mode characteristics, the klystron, because that experiment cannot be done any other way. Otherwise the Gunn bench: it runs from a low voltage supply, there is no vacuum tube to replace, it is considerably safer with students, and every other experiment on the bench is unchanged. Several laboratories own one klystron bench for the mode experiment and Gunn sources for everything else, which is a sensible arrangement.
Almost always the modulation. The meter is a narrow-band amplifier tuned to a fixed audio frequency, so the microwave source must be square-wave modulated at that frequency - check the modulator on the klystron supply or the PIN modulator on a Gunn bench, and check the frequency setting. After that, check that the detector is connected and that the frequency meter has been detuned off resonance, since a meter left on resonance absorbs the power you are trying to measure.
Usually the power level is too high. The meter scale assumes the crystal detector is working in its square-law region, which holds only at low power; above it the scale no longer corresponds to standing wave ratio and the numbers are plausible but incorrect. Reduce the level with the attenuator and repeat. Also check the matched load, because a poor load puts a floor under every reading you can obtain.
Only within the same waveguide band and the same flange standard. Across bands the components will not pass the same frequencies and the flanges will generally not bolt together. If you are extending an existing bench, send us the band and flange details, or a photograph of a flange with a measurement across it, and we will confirm compatibility in writing before quoting.
Yes. The load's own standing wave ratio is the floor on everything the bench can measure: if the load reads 1.05, no component measured on that bench will ever read better, and a well-matched component will look mismatched. Ask for the load's figure, keep it capped and clean, and replace it if it is dropped or damaged. It is a measurement standard in the shape of an accessory.
That depends on your country and the band, so it should be confirmed before purchase rather than after. The equipment radiates, and some administrations regulate operation in these bands even indoors. We will state the operating band in writing so your institution can put the question to the national spectrum authority. We will not tell you that authorisation is unnecessary, because that is not ours to say.
Ray Export is a manufacturer and exporter of laboratory, scientific and educational equipment based at Plot No. 449, HSIIDC Industrial Area, Saha, Ambala, Haryana 133104, India. Enquiries: [email protected], +91-9817319615.
Written by Palak Mall, Senior Content Writer. Technically reviewed by Arvind Kumar, Lab Equipment Specialist, 12+ years in laboratory and testing instrumentation. Last updated: 17 October 2026.
Notes on this listing, published rather than hidden: every product name in this category currently ends with the phrase "For Vocational Training And Didactic...", which is keyword padding rather than product information, and in several records it has overflowed the name field and truncated the part of the name that identifies the item - MTB-0003 loses the waveguide component that distinguishes it. We are stripping the padding and restoring the names. A "Microwave Test Benches" record also appears in our antenna and satellite trainers listing, and klystron and Gunn bench items appear in our microwave lab trainers listing, so the three overlap; we have flagged the structure for correction. Waveguide bands, flange standards, matched load figures and all other specifications will be confirmed in writing on the quotation; verify them before tender submission. Spectrum authorisation requirements differ by country and must be confirmed with your national authority.
Microwave Test Benches are exactness made microwave frameworks, which utilize standard write rectangular wave-manage segments to show the fundamental components of this field for study.
The Microwave Test Benches are exactness made microwave frameworks, which utilize standard compose rectangular wave-control segments to show the basic components of this field for examine. The hardware comprise of : A choice of wave-control parts The power supply for the microwave source A finder A meter, which screens the indicator yield Wall Chart Microwave Components Chart These Training Benches are totally independent and give the way to enable understudies to do pragmatic work at greatly minimal effort. A complete manual containing broad microwave hypothesis and a dynamic arrangement of assignments is provided with the Trainer..
The Microwave Test Benches are exactness made microwave frameworks, which utilize standard compose rectangular wave-manage parts to represent the fundamental components of this field for think about. The hardware comprise of : A determination of wave-direct segments The power supply for the microwave source An indicator A meter, which screens the identifier yield Wall Chart Microwave Components Chart These Training Benches are totally independent and give the way to enable understudies to complete down to earth work at to a great degree minimal effort. A complete manual containing broad microwave hypothesis and a dynamic arrangement of assignments is provided with the Trainer.
The Microwave Test Benches are exactness made microwave frameworks, which utilize standard compose rectangular wave-control segments to delineate the fundamental components of this field for think about. The hardware comprise of : A choice of wave-manage parts The power supply for the microwave source A finder A meter, which screens the indicator yield Wall Chart Microwave Components Chart These Training Benches are totally independent and give the way to enable understudies to do useful work at to a great degree ease. A far reaching manual containing broad microwave hypothesis and a dynamic arrangement of assignments is provided with the Trainer..
This Microwave Test Bench is precision-made microwave systems that use standard type rectangular wave-guide components to illustrate the essential elements of this field for study. These components are available with us in an affordable range and it is highly reliable. We manufacture and trade the high quality components.
Microwave Test Bench that assists the students to understand all the concepts of microwave transmission, various components used for microwave communication and their parameters. 1. Study of Gunn oscillator 2. Frequency, Guide wavelength Measurement 3. VSWR,Reflection Co-efficient Measurement 4. Impedance Measurement