Quick answer. A peel off strength tester measures the force needed to separate two bonded layers, and reports it per unit width - newtons per 25 mm of specimen width, or newtons per millimetre - never as a bare force. It is the standard instrument for adhesive tapes, labels, laminates, coated fabrics, footwear bonds and heat-sealed flexible packaging. The number it produces depends on three things that are part of the method and not of the material: the peel angle, the rate of peel, and the portion of the trace you average. This category holds one product, PLC-0001, and the category name does not describe it - see the note below.
A note on this category, published rather than hidden. This category is currently named "PLC Testing Lab Trainers" and its product code prefix is PLC-, but it contains exactly one product and that product is a peel off strength tester - an adhesion testing instrument with no connection to programmable logic controllers. The prefix was evidently generated from the category name. We have flagged it for correction: the category should be renamed for adhesion testing, or the instrument moved into the materials testing range and this address redirected. If you were looking for PLC trainers, they are in our electronics lab trainer kits and mechatronics ranges, not here. If you were looking for peel strength testing, you are in the right place and the rest of this page is for you.
This is the single most common mistake in peel testing, and it causes more disputes between a supplier and a customer than any instrument fault. A peel result is not a property of the bond alone. The same bond, tested honestly in two different geometries, gives two different numbers.
| Geometry | What happens | Typically used for |
|---|---|---|
| 180 degree peel | The flexible layer is folded back on itself and pulled in the opposite direction to the bond line. | Adhesive tapes and labels on a rigid substrate; the most widely specified tape test. |
| 90 degree peel | The flexible layer is pulled perpendicular to the substrate, which has to be held in a moving or rolling fixture to keep the angle constant. | Coated fabrics, laminates to rigid panels, and flooring or wall coverings. |
| T-peel | Both layers are flexible, so both are gripped and pulled apart; neither is supported. | Flexible packaging seals, film laminates, and fabric-to-fabric bonds. |
So the first question on an enquiry is not what capacity you need, it is which geometry your customers specify. The geometry decides the fixture, and the fixture is the part of the instrument that does the work. A tester quoted without naming the fixture is not quoted for your test.
Peel force scales with specimen width, which is why results are reported per unit width. The practical consequence is that your specimen cutter is as much a part of the measurement as the load cell.
If the method calls for 25 mm specimens and your strips are cut at 23 mm with scissors, and the result is divided by the nominal 25 mm, every figure you report is low by about eight per cent - consistently, invisibly, and in the direction that understates your product. The remedies are ordinary: cut with a parallel-blade specimen cutter rather than by hand, measure the actual width of each specimen and divide by what you measured, and discard specimens with a nicked or ragged edge, because a peel front propagates from a flaw.
Adhesives are viscoelastic: their response depends on how fast you load them. Pulling faster generally gives a higher peel force, and the difference is not small. The rate of peel is specified in every peel method, and a result reported without the rate cannot be compared with anything.
Two things follow for the specification of the instrument. The crosshead speed has to cover the rate your method names, with the resolution to set it exactly rather than approximately. And the speed has to be stable under load, because a drive that slows as the force rises is changing the test condition during the test.
A peel test rarely produces a steady force. The trace typically shows an initial peak as the bond starts to separate, then an irregular plateau as the peel front travels, sometimes with the sawtooth pattern that indicates stick-slip separation. Peel methods deal with this by averaging the force over a defined portion of the peel length and discarding the start-up region.
This is why an instrument that displays only a peak value is the wrong instrument for peel work, however well it is made. Ask for:
And record the mode of failure alongside the number, by eye: adhesive failure at the interface, cohesive failure within the adhesive, or substrate failure. A high peel figure obtained by tearing the substrate is not a measure of the bond at all, and a report without the failure mode conceals that.
Two periods have to elapse before a peel specimen is ready, and laboratories under delivery pressure shorten both.
If your laboratory cannot hold a standard atmosphere yet, record the actual temperature and humidity on every peel report. A number with its conditions attached is usable. A number without them is not.
We would rather say this than sell the wrong thing. A peel test is a tensile test with a particular fixture. If your laboratory already has a universal testing machine with a low-force load cell, adequate crosshead speed control and trace export, a peel fixture and a specimen cutter may be all you need.
The dedicated tester earns its place where any of the following is true:
Tell us which case you are in. If the answer is a fixture and a cutter, we would rather quote a fixture and a cutter.
Adhesion and bond testing appears across several of our ranges: coated fabric testing equipment for laminate and coating adhesion, footwear testing equipment for sole and upper bond strength, paper and packaging testing equipment for seal and ply bond, fabric testing equipment for textile tensile work, and metal testing machines for the universal and tensile machines a peel fixture mounts to. The full testing machines category lists every testing range we manufacture.
Check four things in this order: the peel angle, the rate of peel, the specimen width and how it was measured, and the portion of the trace averaged. A difference in any one of them is enough to explain a sizeable gap, and the angle alone can change the figure substantially. Exchange the full test conditions with your customer before exchanging accusations about the material.
Per unit width, and the width must be the one you measured rather than the one you intended. Reporting a bare force in newtons is only meaningful if the specimen width is stated in the same report, and even then it invites arithmetic errors downstream. Report newtons per 25 mm or newtons per millimetre, and state the width tested.
Unstable, stick-slip peeling: the peel front arrests and then jumps. It is common with stiff backings and brittle adhesives, and it matters because the average force depends on how much of the oscillation your averaging window captures. If you see a sawtooth, record it, keep the trace, and be cautious about comparing the average with a result from a different instrument whose sampling rate may have smoothed it away.
The load frame and the control system can. Each geometry needs its own fixture: T-peel needs only two grips, 180 degree needs a backing plate, and 90 degree needs a fixture that moves the substrate laterally so the angle stays at 90 degrees as the peel front travels. Specify the geometries you need at the quotation stage, because retrofitting a rolling 90 degree fixture later is more expensive than ordering it with the machine.
Usually the opposite. Peel forces are small, and an oversized load cell cannot resolve them: a cell sized for several kilonewtons will not give a usable reading on a tape peeling at a few newtons. Size the cell for resolution across your working range, and if your forces span a wide range, ask about an interchangeable cell rather than one cell that compromises both ends.
Tell us what your tender or quality system requires and we will state in writing what accompanies the model quoted. All specifications on this page are generic descriptions of the method rather than claims about a particular model; the figures for the instrument we quote will be given in writing. If you need traceable calibration from an accredited laboratory, raise it at the enquiry stage, because it is arranged separately and affects price and lead time.
Ray Export is a manufacturer and exporter of laboratory, scientific and educational equipment based at Plot No. 449, HSIIDC Industrial Area, Saha, Ambala, Haryana 133104, India. Enquiries: [email protected], +91-9817319615.
Written by Palak Mall, Senior Content Writer. Technically reviewed by Arvind Kumar, Lab Equipment Specialist, 12+ years in laboratory and testing instrumentation. Last updated: 17 October 2026.
This page is deliberately shorter than our other category pages, because the category holds one product and padding it would serve nobody. The category name and product code prefix are wrong for the contents, as described above, and are flagged for correction. Method descriptions here are generic; the applicable standard differs by material and market, and we will confirm the standard an instrument is built to, in writing, at the quotation stage. Specifications are confirmed in writing on the quotation; for certifications, contact us for the latest copies of the relevant certificates before tender submission.
Peel Off Strength Tester - We offer Peel Off Strength Tester, to test the self-adhesiveness of froth sheet. The quality of bond of self-glue froth sheets to the substrate is controlled by sticking a rectangular test example cut out of the sheet under test over a smooth glass surface, squeezing it down with the assistance of elastic lined metal roller, and peeling off one of its closures from the substrate at an edge of 180° at a steady speed with the assistance of a mechanized course of action. and The power expected to peel off the test example gives a measure of the peel quality of the froth sheet under test. common lab types of gear producers.