Peel Off Strength Tester

Quick answer. A peel off strength tester measures the force needed to separate two bonded layers, and reports it per unit width - newtons per 25 mm of specimen width, or newtons per millimetre - never as a bare force. It is the standard instrument for adhesive tapes, labels, laminates, coated fabrics, footwear bonds and heat-sealed flexible packaging. The number it produces depends on three things that are part of the method and not of the material: the peel angle, the rate of peel, and the portion of the trace you average. This category holds one product, PLC-0001, and the category name does not describe it - see the note below.

A note on this category, published rather than hidden. This category is currently named "PLC Testing Lab Trainers" and its product code prefix is PLC-, but it contains exactly one product and that product is a peel off strength tester - an adhesion testing instrument with no connection to programmable logic controllers. The prefix was evidently generated from the category name. We have flagged it for correction: the category should be renamed for adhesion testing, or the instrument moved into the materials testing range and this address redirected. If you were looking for PLC trainers, they are in our electronics lab trainer kits and mechatronics ranges, not here. If you were looking for peel strength testing, you are in the right place and the rest of this page is for you.

The angle is part of the number

This is the single most common mistake in peel testing, and it causes more disputes between a supplier and a customer than any instrument fault. A peel result is not a property of the bond alone. The same bond, tested honestly in two different geometries, gives two different numbers.

GeometryWhat happensTypically used for
180 degree peelThe flexible layer is folded back on itself and pulled in the opposite direction to the bond line.Adhesive tapes and labels on a rigid substrate; the most widely specified tape test.
90 degree peelThe flexible layer is pulled perpendicular to the substrate, which has to be held in a moving or rolling fixture to keep the angle constant.Coated fabrics, laminates to rigid panels, and flooring or wall coverings.
T-peelBoth layers are flexible, so both are gripped and pulled apart; neither is supported.Flexible packaging seals, film laminates, and fabric-to-fabric bonds.

So the first question on an enquiry is not what capacity you need, it is which geometry your customers specify. The geometry decides the fixture, and the fixture is the part of the instrument that does the work. A tester quoted without naming the fixture is not quoted for your test.

Per unit width, and why specimen cutting is half the result

Peel force scales with specimen width, which is why results are reported per unit width. The practical consequence is that your specimen cutter is as much a part of the measurement as the load cell.

If the method calls for 25 mm specimens and your strips are cut at 23 mm with scissors, and the result is divided by the nominal 25 mm, every figure you report is low by about eight per cent - consistently, invisibly, and in the direction that understates your product. The remedies are ordinary: cut with a parallel-blade specimen cutter rather than by hand, measure the actual width of each specimen and divide by what you measured, and discard specimens with a nicked or ragged edge, because a peel front propagates from a flaw.

Rate of peel, and why adhesives are not like metals

Adhesives are viscoelastic: their response depends on how fast you load them. Pulling faster generally gives a higher peel force, and the difference is not small. The rate of peel is specified in every peel method, and a result reported without the rate cannot be compared with anything.

Two things follow for the specification of the instrument. The crosshead speed has to cover the rate your method names, with the resolution to set it exactly rather than approximately. And the speed has to be stable under load, because a drive that slows as the force rises is changing the test condition during the test.

Read the trace, not the peak

A peel test rarely produces a steady force. The trace typically shows an initial peak as the bond starts to separate, then an irregular plateau as the peel front travels, sometimes with the sawtooth pattern that indicates stick-slip separation. Peel methods deal with this by averaging the force over a defined portion of the peel length and discarding the start-up region.

This is why an instrument that displays only a peak value is the wrong instrument for peel work, however well it is made. Ask for:

  • Averaging over a user-defined peel length, with the start-up region excluded as the method requires.
  • A force-against-displacement trace that can be exported, because the shape of the trace is diagnostic. A clean plateau, a sawtooth and a sudden drop to zero mean three different things about the bond.
  • A sampling rate high enough not to smooth away the sawtooth, which is the feature that tells you the failure is unstable.

And record the mode of failure alongside the number, by eye: adhesive failure at the interface, cohesive failure within the adhesive, or substrate failure. A high peel figure obtained by tearing the substrate is not a measure of the bond at all, and a report without the failure mode conceals that.

Conditioning and dwell: the two waiting periods

Two periods have to elapse before a peel specimen is ready, and laboratories under delivery pressure shorten both.

  • Dwell after bonding. An adhesive bond continues to develop after it is made. Peel methods specify a residence time between applying the specimen and testing it, and a bond tested after ten minutes is not the bond tested after twenty-four hours. If your customer specifies a dwell, it is not padding.
  • Conditioning in a standard atmosphere. Adhesive behaviour moves with temperature and humidity, and so does the stiffness of the backing. The method specifies the atmosphere; a laboratory testing in whatever the room happens to be is adding a seasonal variable to its results.

If your laboratory cannot hold a standard atmosphere yet, record the actual temperature and humidity on every peel report. A number with its conditions attached is usable. A number without them is not.

Before you buy: you may already own most of this instrument

We would rather say this than sell the wrong thing. A peel test is a tensile test with a particular fixture. If your laboratory already has a universal testing machine with a low-force load cell, adequate crosshead speed control and trace export, a peel fixture and a specimen cutter may be all you need.

The dedicated tester earns its place where any of the following is true:

  • Peel is a routine, high-volume test and tying up the universal machine is costly.
  • Your forces are low enough that the universal machine load cell cannot resolve them. This is the usual reason: a load cell sized for metals will not measure tape peel usefully, and resolution at the bottom of the range, not maximum capacity, is the specification that matters here.
  • You need a 90 degree rolling fixture that your machine cannot accommodate.

Tell us which case you are in. If the answer is a fixture and a cutter, we would rather quote a fixture and a cutter.

What to tell us when you enquire

  • The geometry your customers specify - 180 degree, 90 degree or T-peel - and the standard it is named in.
  • The expected force range and the specimen width, so the load cell is sized for resolution at your forces rather than for a headline capacity.
  • The rate of peel the method requires.
  • What the specimens are: tape, label, laminate, coated fabric, footwear bond or heat seal. The grips and the backing plate differ.
  • Whether you need trace export and averaging over a defined length, and into what software.
  • Supply voltage, frequency and destination country.

Adhesion and bond testing appears across several of our ranges: coated fabric testing equipment for laminate and coating adhesion, footwear testing equipment for sole and upper bond strength, paper and packaging testing equipment for seal and ply bond, fabric testing equipment for textile tensile work, and metal testing machines for the universal and tensile machines a peel fixture mounts to. The full testing machines category lists every testing range we manufacture.

Frequently asked questions

Why is my peel result different from my customer's on the same material?

Check four things in this order: the peel angle, the rate of peel, the specimen width and how it was measured, and the portion of the trace averaged. A difference in any one of them is enough to explain a sizeable gap, and the angle alone can change the figure substantially. Exchange the full test conditions with your customer before exchanging accusations about the material.

Should peel strength be reported as a force or a force per width?

Per unit width, and the width must be the one you measured rather than the one you intended. Reporting a bare force in newtons is only meaningful if the specimen width is stated in the same report, and even then it invites arithmetic errors downstream. Report newtons per 25 mm or newtons per millimetre, and state the width tested.

What does a sawtooth trace mean?

Unstable, stick-slip peeling: the peel front arrests and then jumps. It is common with stiff backings and brittle adhesives, and it matters because the average force depends on how much of the oscillation your averaging window captures. If you see a sawtooth, record it, keep the trace, and be cautious about comparing the average with a result from a different instrument whose sampling rate may have smoothed it away.

Can one instrument do 180 degree, 90 degree and T-peel?

The load frame and the control system can. Each geometry needs its own fixture: T-peel needs only two grips, 180 degree needs a backing plate, and 90 degree needs a fixture that moves the substrate laterally so the angle stays at 90 degrees as the peel front travels. Specify the geometries you need at the quotation stage, because retrofitting a rolling 90 degree fixture later is more expensive than ordering it with the machine.

Does the load cell capacity need to be high?

Usually the opposite. Peel forces are small, and an oversized load cell cannot resolve them: a cell sized for several kilonewtons will not give a usable reading on a tape peeling at a few newtons. Size the cell for resolution across your working range, and if your forces span a wide range, ask about an interchangeable cell rather than one cell that compromises both ends.

Is a calibration certificate supplied?

Tell us what your tender or quality system requires and we will state in writing what accompanies the model quoted. All specifications on this page are generic descriptions of the method rather than claims about a particular model; the figures for the instrument we quote will be given in writing. If you need traceable calibration from an accredited laboratory, raise it at the enquiry stage, because it is arranged separately and affects price and lead time.

Key takeaways

  • The geometry is part of the result. 180 degree, 90 degree and T-peel give different numbers on the same bond. Specify the geometry before the capacity.
  • Report per unit width, using the width you measured. A specimen cut 2 mm narrow shifts every result by about eight per cent.
  • Rate of peel is a test condition, not a convenience. Adhesives are rate-dependent; a result without the rate is not comparable.
  • Average over a defined length and keep the trace. A peak-only readout is the wrong instrument for peel, and the trace shape is diagnostic.
  • Record the failure mode. A high figure obtained by tearing the substrate is not a measure of the bond.
  • Resolution beats capacity. Size the load cell for the small forces you actually measure, and check whether a fixture on your existing universal machine would do the job.

About this page

Ray Export is a manufacturer and exporter of laboratory, scientific and educational equipment based at Plot No. 449, HSIIDC Industrial Area, Saha, Ambala, Haryana 133104, India. Enquiries: [email protected], +91-9817319615.

Written by Palak Mall, Senior Content Writer. Technically reviewed by Arvind Kumar, Lab Equipment Specialist, 12+ years in laboratory and testing instrumentation. Last updated: 17 October 2026.

This page is deliberately shorter than our other category pages, because the category holds one product and padding it would serve nobody. The category name and product code prefix are wrong for the contents, as described above, and are flagged for correction. Method descriptions here are generic; the applicable standard differs by material and market, and we will confirm the standard an instrument is built to, in writing, at the quotation stage. Specifications are confirmed in writing on the quotation; for certifications, contact us for the latest copies of the relevant certificates before tender submission.

Peel Off Strength Tester

PLC-0001

Peel Off Strength Tester - We offer Peel Off Strength Tester, to test the self-adhesiveness of froth sheet. The quality of bond of self-glue froth sheets to the substrate is controlled by sticking a rectangular test example cut out of the sheet under test over a smooth glass surface, squeezing it down with the assistance of elastic lined metal roller, and peeling off one of its closures from the substrate at an edge of 180° at a steady speed with the assistance of a mechanized course of action. and The power expected to peel off the test example gives a measure of the peel quality of the froth sheet under test. common lab types of gear producers.

Supply of PLC Testing Lab Trainers from India to Finland, France, Tanzania, Thailand, Sri Lanka, Congo, Costa Rica, Ivory Coast, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Djibouti, Dominica, Afghanistan, Albania, Algeria, Andorra, Angola, Antigua and Barbuda, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Bangladesh, Barbados, United Arab Emirates (UAE, Dubai), United Kingdom (London), United States, Uruguay, Uzbekistan, Vanuatu, Belarus, Belgium, Belize, Benin, Bhutan, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei, Bulgaria, Burkina Faso, Burma Myanmar, East Timor, Ecuador, Egypt, El Salvador, Equatorial Guinea, Eritrea, Estonia, Ethiopia (Addis Ababa), Fiji, Burundi, Cambodia, Cameroon, Canada, Cape Verde, Central African Republic, Chad, Chile, Colombia. Math Teaching Kit suppliers in Luxembourg, Macedonia, Madagascar, Malawi (Lilongwe), Malaysia (Kuala Lumpur), Maldives, Mali, Malta, Marshall Islands, Mauritania, Mexico, Micronesia, Moldova, Monaco, Mongolia, Montenegro, Morocco, Mozambique, Namibia, Comoros, Congo. Math Teaching Kit manufacturers in Dominican Republic, Gabon, Gambia, Georgia, Germany, Ghana, Greece, Grenada, Guatemala, Guinea, Guinea-Bissau, Guyana, Haiti, Honduras, Hungary, Iceland, Indonesia, Iran, Iraq, Niger, Nigeria (Abuja), Norway, Oman, Palau, Panama, Papua New Guinea, Paraguay, Peru, Philippines (Manila), Poland, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya (Nairobi), Kiribati, Korea, Kuwait, Kyrgyzstan, Laos, Latvia, Lebanon, Lesotho, Liberia, Liechtenstein, Lithuania, Mauritius, Nauru, Nepal, Netherlands, New Zealand, Nicaragua, Portugal, Qatar, Romania, Russia, Rwanda (Kigali). Math Teaching Kit exportets to Saint Kitts and Nevis, Saint Lucia, Saint Vincent and the Grenadines, Samoa, San Marino, Sao Tome and Principe, Saudi Arabia, Senegal, Serbia, Seychelles, Sierra Leone, Singapore, Slovakia, Slovenia, Solomon Islands, Somalia, South Africa, South Sudan, Spain, Sudan, Suriname, Swaziland, Sweden, Switzerland, Syria, Tajikistan, Togo, Tonga, Trinidad and Tobago, Tunisia, Turkey, Turkmenistan, Tuvalu, Uganda (Kampala), Ukraine, Venezuela, Vietnam, Yemen , Zambia (Lusaka), Zimbabwe