Quick answer. Settle the hardness scale your specification names before comparing the eight hardness instruments. Shore A suits soft and medium elastomers, Shore D hard rubber, and IRHD is used where specifications call for it, and readings do not convert reliably between them. Then decide stand and dwell: a stand applies a defined force square to the surface, and a timed reading controls the creep that makes rubber hardness fall second by second. For seals and gaskets, though, compression set, recovery and tension set predict service life far better than hardness, and they depend on the multi-cell ageing oven RTE-0025.
Twenty-five instruments, coded RTE-0001 to RTE-0025, for a rubber and elastomer laboratory. Eight of them measure hardness, which tells you something about how rubber is specified. The rest cover compression - compression set, compressibility and recovery and three compression testing machines; permanent deformation in tension; density by specific gravity; DIN abrasion; flexing and fatigue with De-Mattia and Ross flex testers; impact resilience; split tear; rub proofness; gas permeability; a specimen cutting press; and a multi-cell ageing oven.
Eight hardness instruments is not redundancy. Rubber hardness is measured on several scales with several instrument types, the scales are not interchangeable, and a specification names one of them. The section below is the part of this page worth reading before you compare prices, because getting it wrong produces numbers that look perfectly reasonable and that your customer cannot use.
Three things make a rubber hardness reading reproducible. Miss any one and two laboratories testing the same compound will disagree.
| What to settle | Why | |
|---|---|---|
| Scale | Shore A, Shore D or IRHD - whichever your specification names | Shore A suits soft and medium elastomers, Shore D hard rubber and rigid plastics, and IRHD is used where specifications call for it and in a micro form for thin or small specimens. Conversion between the scales is approximate at best; a number on the wrong scale is not convertible into a compliant result |
| Stand | Hand-held or stand-mounted with a defined mass | A hand-held durometer is pressed by a person, so the applied force and the squareness of contact vary with the operator. A stand applies a defined force perpendicular to the surface, and that is what turns a reading into a measurement. If results go on a certificate, buy the stand |
| Dwell | The time between contact and reading, as the standard specifies it | Rubber creeps under the indenter, so the indicated hardness falls as the seconds pass. Standards specify a dwell - one second, three seconds, fifteen seconds depending on the method - and a hardness value quoted without its dwell time is incomplete. This is what the dual-needle and timed digital instruments exist to control |
Two further points that decide whether a hardness number is valid at all. Specimen thickness: the methods specify a minimum, and a specimen thinner than the minimum reads harder because the indenter feels the bench through it; stacking thin pieces is permitted only where the method says so, and on moulded products it usually is not. Position: readings must be taken a stated distance from any edge and from each other, and the median of several readings is reported rather than one. Ask us which standard each of the eight instruments implements, whether a stand and a timer are included, and what reference rubber blocks are supplied for checking - a durometer is checked against reference blocks, and a laboratory without them has no way of knowing its instrument has drifted.
For anyone making gaskets, O-rings, bushes or mountings, the instruments that matter most are not the hardness ones. They are compression set, compressibility and recovery, and tension set, and they all measure the same underlying thing: how much deformation is permanent after the rubber has been held deformed, hot, for a long time.
That is the property behind almost every leak in service. A compound that recovers fully at twenty-five degrees may take a permanent set at a hundred, and a seal that has taken a set no longer presses on the flange. Ask about: the jigs and spacers, and whether the compression is to a fixed deflection or a fixed load, because the two methods are different and give different numbers; the parallelism and flatness of the platens, since a non-parallel jig loads the specimen unevenly; the temperature and time the method specifies and how you will hold them, which usually means the ageing oven; and the recovery period and temperature before measurement, which is part of the method and not a convenience.
The multi-cell ageing oven, RTE-0025, is worth more attention than its position at the end of the listing suggests. It is not an ordinary laboratory oven, and the reason is specific: antidegradants, plasticisers and other volatiles migrate out of one compound and into another when samples share an air space, which changes the ageing behaviour of both. The standards therefore require that different compounds are aged in separate cells with their own air flow. A single-chamber oven full of mixed compounds produces ageing results that are a property of the batch of samples rather than of any one compound. Ask how many cells there are, whether each has its own air change and at what rate, the temperature range and uniformity, and how temperature is recorded, because an ageing test unrecorded for seventy-two hours is an act of faith.
The DIN abrasion tester reports volume loss when a specimen is run against an abrasive sheet under load. The subtlety is the sheet: abrasive sheets vary from batch to batch, so the method requires the sheet to be checked against a reference compound and the result expressed relative to it. Ask what reference rubber is supplied, where it is obtained, and what sheets cost - as everywhere in materials testing, the consumable is the running cost and in this case it is also the calibration.
The De-Mattia and Ross flex testers find cracking under repeated bending, which is how rubber fails in service far more often than by breaking. Ask the number of specimens per run, the stroke and frequency, whether a cold chamber is available, and how cracks are assessed and graded. Note that a De-Mattia tester also appears in the fabric testing equipment listing, spelled slightly differently; they are the same kind of instrument applied to different materials, so quote by code and full name.
The split tear tester and the specimen cutting press belong together, because tear results are more sensitive to specimen preparation than any other rubber test. A die that is blunt, or a cut edge with a small nick in it, starts a tear early and scatters the results - and the scatter looks like material variation rather than preparation. Budget for dies, for resharpening and for the discipline of replacing them. Ask which dies each test needs, whether the press takes them all, and what resharpening costs locally. Tensile and tear strength themselves are measured on a machine from the tensile testing machines or universal testing machines range, and for rubber you want a low-capacity load cell and grips that do not cut the specimen.
Two specific gravity instruments are listed, and density is the most undervalued routine test in a rubber laboratory. It is quick, it needs no consumables beyond water, and a change in density between batches is the earliest warning that the filler loading has changed - which usually means the compound has changed without anyone saying so. For a goods-inwards check on bought-in compound it is the first test to run. Ask about the balance resolution, the hydrostatic arrangement and whether temperature correction is applied.
Gas permeability matters if your products hold gas - bladders, diaphragms, inflatable seals, pneumatic components. Ask which gas the test uses, what pressure difference is applied, the specimen area and thickness required, how long a determination takes and what gas supply is needed at the position. Ask also what the instrument reports, since permeability, transmission rate and diffusion coefficient are related but not the same quantity, and a specification names one.
The names in this listing overlap more than they should. Eight records are hardness testers, including one whose name ends in a digit because it duplicates another; three are compression testing machines; and two are specific gravity instruments - and the listing does not say what separates them. One record spells a standards body as an ordinary word rather than an acronym. So: ask for the specification sheets of the records you are choosing between, side by side, and compare on scale, stand, dwell control and whether reference blocks are included. And as always on this part of the catalogue, quote the code and the full product name together and have both confirmed on the proforma invoice. If a name looks like a duplicate, ask; it is more likely to be our catalogue data than a second product.
For each hardness instrument: scale, standard implemented, whether a test stand and timer are included, dwell control, reference blocks supplied, and calibration certificate. For the set and recovery tests: jigs, spacers, fixed-deflection or fixed-load method, platen flatness, and the oven they depend on. For the ageing oven: number of cells, air change per cell, temperature range, uniformity and recording. For abrasion: reference compound supplied, sheet source and cost. For the flex testers: specimens per run, stroke, frequency, cold capability and assessment method. For cutting: dies included, resharpening arrangement. For all: calibration and traceability, bench space and mass, services at the position, nominal voltage, frequency and plug pattern, consumables and spares with prices and lead times, and the warranty and local service position. For any certification stated here, contact us for the latest copies of the relevant certificates before you rely on it in a tender or an accreditation file.
Related ranges: foam testing equipment for cellular materials, plastic testing equipment for thermoplastics and films, footwear testing equipment for soling and shoe components, and coated fabric testing equipment for rubber-proofed cloth. The full set is listed under raw products testing. For a quotation, tell us the destination country, the standards your customers name, which hardness scale your specifications use and what you make - a gasket plant, a tyre retreader and a footwear soling unit need three different subsets of this page.
The one your specification names. Shore A suits soft and medium elastomers, Shore D hard rubber and rigid plastics, and IRHD is used where specifications call for it, including a micro form for thin or small specimens. Conversion between scales is approximate at best, so a number on the wrong scale cannot be turned into a compliant result. Then choose hand-held or stand-mounted: a hand-held durometer varies with the operator's force and squareness, while a stand applies a defined force perpendicular to the surface. If results are reported to customers, buy the stand, and ask what reference rubber blocks come for checking drift.
Usually dwell, thickness or position. Rubber creeps under the indenter, so the indicated hardness falls as seconds pass; methods specify a dwell of one, three or fifteen seconds, and a value quoted without its dwell is incomplete, which is what dual-needle and timed digital instruments control. A specimen thinner than the method's minimum reads harder because the indenter feels the bench through it, and stacking thin pieces is allowed only where the method says so, rarely on moulded products. Readings must also be a stated distance from edges and from each other, with the median of several reported rather than a single reading.
Compression set, compressibility and recovery, and tension set. They all measure how much deformation stays permanent after rubber has been held deformed, hot, for a long time, which is the property behind almost every leak in service. A compound that recovers fully at 25 deg C may take a permanent set at 100 deg C, and a seal that has taken a set no longer presses on the flange. Ask about the jigs and spacers, whether compression is to a fixed deflection or a fixed load, since the two give different numbers, the flatness and parallelism of the platens, and the recovery period before measurement.
Because compounds contaminate each other. Antidegradants, plasticisers and other volatiles migrate out of one compound and into another when samples share an air space, changing the ageing behaviour of both. The methods therefore require different compounds to be aged in separate cells with their own air flow, and a single-chamber oven full of mixed compounds produces results that belong to the batch of samples rather than to any one compound. Ask how many cells there are, whether each has its own air change and at what rate, the temperature range and uniformity, and how temperature is recorded across a long test.
Consumables and specimen preparation. The abrasion tester reports volume loss against an abrasive sheet, and sheets vary from batch to batch, so results are expressed relative to a reference compound; ask what reference rubber is supplied and where sheets come from, since the consumable is also the calibration. For the De-Mattia and Ross flex testers, ask the specimens per run, stroke, frequency, cold capability and how cracks are graded. Tear results are the most sensitive to preparation: a blunt die or a nicked edge starts a tear early, so ask which dies the cutting press takes and how resharpening is handled.
Specific gravity is the most undervalued routine test in a rubber laboratory. It is quick, needs nothing beyond water, and a density change between batches is the earliest warning that the filler loading, and so the compound, has changed; it is the first goods-inwards check on bought-in compound. Ask the balance resolution, the hydrostatic arrangement and whether temperature correction is applied. Gas permeability matters if your products hold gas, such as bladders, diaphragms and inflatable seals. Ask which gas and pressure difference are used, the specimen size, the test time, and whether it reports permeability, transmission rate or diffusion coefficient.
Ray Export is a manufacturer and exporter of laboratory, scientific and educational equipment based at Plot No. 449, HSIIDC Industrial Area, Saha, Ambala, Haryana 133104, India. Enquiries: [email protected], +91-9817319615.
Written by Palak Mall, Senior Content Writer. Technically reviewed by Arvind Kumar, Lab Equipment Specialist, 12+ years in laboratory and testing instrumentation. Last updated: 17 October 2026.
Specifications for RTE-0001 to RTE-0025, including hardness scale and dwell control, test stand and reference blocks, jig and platen details, ageing oven cells and air change, and the dies supplied, are confirmed in writing on the quotation. For calibration or traceability documents, contact us for the latest copies of the relevant certificates. No price is quoted on this page. Listing note: the eight hardness records include one whose name ends in a digit because it duplicates another; three compression testing machines and two specific gravity instruments are listed with nothing to separate them; one record spells a standards body as an ordinary word; and a De-Mattia tester also appears in the fabric testing listing under a slightly different spelling.
Split Tear Tester - Micro cell sheets utilized as a part of the fabricate of hawai chappals and shoe soles ought to be impervious to slitting on the off chance that they get split over their thickness. and The capacity of miniaturized scale cell elastic sheets and soles to oppose proliferation of parts is dictated by part a rectangular test example taken from the sheet or sole under test over its thickness and pulling the two split closures from each other. The most extreme load applied before disappointment gives the measure of the obstruction of the sheet or sole under test to part. and The Split Tear Tester for microcellular sheets and shoe soles comprises of a course of action to hold two split tongues of the test example in two separate grasps in a vertical plane, a game plan to move one of the grasps descending at a predetermined speed, and a game plan to record the greatest load applied while spreading the split. common lab types of gear makers.
Contact Ray Export for your School, College Civil and Mechanical Engineering Lab Instruments. We are the best engineering educational equipments india, engineering equipments distributors, engineering lab instruments, engineering equipments manufacturer, thermodynamics engineering equipment, engineering educational lab equipments in Ambala, India.
Ross Flex Tester - The capacity of elastic vulcanizates to oppose split development of a punctured example when subjected to rehashed bowing can be assessed with the assistance of a Ross Flex Tester. The& Ross Flex Tester comprises of a course of action to hold one end of the test examples in a wavering grasp to such an extent that they can be twisted through an edge of 90° at the predetermined recurrence, and customizable guide rollers to keep the free end of the test examples pushed down amid the bowing tasks. and The example grasp has arrangement for holding six sets of test examples at once. The grasp is turned on two metal rollers to give a smooth low-grinding development. Development of the grasp causes test examples to twist around a silver steel ground bar with its hub matching with the hub of development of the hold. The development is given by an engine and V-belts through an unpredictable and connection system. An electronic computerized counter with reinforcement memory is given to record the aggregate quantities of twisting cycles..
Contact Ray Export for your School, College Civil and Mechanical Engineering Lab Instruments. We are the best engineering lab equipments, engineering equipment manufacturer india, engineering equipment india, chemical engineering lab equipments, control engineering laboratory equipments, control engineering laboratory equipment, control engineering lab equipment in Ambala, India.
Rub Proofness Tester - The Rub Proofness Tester is utilized for testing the speed of shade of light cowhides to wet and dry rubbing and estimating the exchange of shading to the material with which it is rubbed. The test piece is rubbed with a pivoting roundabout felt cushion squeezed against it under the predetermined load, and the quantity of insurgencies of the cushion required to create certain impacts is estimated. and This hardware can likewise be utilized to survey the level of harm to a material amid gentle wet scraped area to mirror the scraping activity of a sodden hose on footwear insole amid wear. A pivoting wet round felt cushion, secured with a worsted repp texture, rubs the test example under a predetermined contact constrain. The general harm to the test examples is surveyed by estimating their mass misfortune after a predefined quantities of revolutions.&&&&&& Civil lab types of gear makers.
Contact Ray Export for your School, College Civil and Mechanical Engineering Lab Instruments. We are the best engineering lab equipments, engineering equipment manufacturer india, engineering equipment india, chemical engineering lab equipments, control engineering laboratory equipments, control engineering laboratory equipment, control engineering lab equipment in Ambala, India.
Shore A Hardness Tester - Hardness is a standout amongst the most vital properties which decide the reasonableness of any elastic segment for its planned end utilize. Wrong hardness may nullify the fundamental point for which the segment is composed. Along these lines exact and dependable assurance of hardness is of outrageous significance in elastic field. and The Shore A Hardness Tester, likewise called as Durometer, is the most usually utilized instrument for estimation of hardness of any elastic segment. It depends on the estimation of profundity of space of an indentor of determined shape being constrained into the material under a predetermined weight. This test technique is an experimental strategy proposed fundamentally for control purposes. No basic relationship exists between hardness controlled by this test strategy and any central property of the material being tried. and The durometer comprises basically of the accompanying segments: A weight foot having a gap of 2.5 to 3.2 mm breadth focused no less than 6 mm from any edge of the foot, and An indentor made of solidified steel having a width in the vicinity of 1.15 and 1.40 mm, with a tip fit as a fiddle of a truncated cone of 35° included edge and a level surface of 0.79 mm measurement, and A dial on which the measure of augmentation of the tip of indentor might be perused as far as graduations running from zero for full expansion of 2.46 to 2.54 mm, to 100 for zero expansion, acquired by setting the weight foot and indentor in firm contact with a level bit of glass. and An aligned spring for applying the predefined power to the indentor. common lab types of gear producers. Elastic Testing Lab hardware manufacturers ,exporters and Rubber Testing Lab providers
Contact Ray Export for your School, College Civil and Mechanical Engineering Lab Instruments. We are the best engineering educational equipments india, engineering equipments distributors, engineering lab instruments, engineering equipments manufacturer, thermodynamics engineering equipment, engineering educational lab equipments in Ambala, India.
Multi-Cell Aging Oven - Accelerated maturing test is directed on polymers to assess their protection from decay with the progression of time. For this reason the test example is subjected to controlled breaking down impacts at raised temperatures for clear periods, after which the impact of such effects on their properties are resolved and thought about against the comparing properties of un-matured polymers. One of the device utilized for this object is the multi-cell maturing broiler. The& Multi-Cell Aging Oven is intended to do quickened maturing on polymers by subjecting them to crumbling by controlled dissemination of air at a raised temperature under climatic weight. It comprises of an aluminum square having seven gaps of 100 mm width, each opening speaking to a cell. The test examples are unreservedly suspended inside these cells on wire suspenders for maturing. and The square is warmed with the assistance of cartridge electric radiators set in it. Its temperature is controlled with the assistance of a computerized temperature controller-cum-pointer. The square is set inside a sheet metal bureau and is encompassed with mineral fleece for giving it warm protection. common lab types of gear makers.
Contact Ray Export for your School, College Civil and Mechanical Engineering Lab Instruments. We are the best manufacturer india, engineering equipment india, chemical engineering lab equipments, control engineering laboratory equipments, control engineering laboratory equipment, control engineering lab equipment in Ambala, India.